BACKGROUND

In a pediatric setting, the need for lifetime oral anticoagulation is increasing because of currency of extracardiac total cavo-pulmonary connection (TCPC) and pediatric valve surgery. We evaluated a new compact device "CoaguChek XS" for measuring prothrombin time-internatinal normalized ratio (PT-INR).

METHODS

The international normalized ratio (INR) values obtained from 71 patients (223 samples) by a CoaguChek XS were compared with those obtained by a laboratory-based coagulation analyzer.

RESULTS

The values from the CoaguChek XS had a significant correlation with the laboratory based results. (r2 = 0.92, p < 0.01, regression line y = 1.05 x -0.02).

CONCLUSION

The CoaguChek XS will be useful in pediatric management.