研究光子和碳离子辐照后 Presage™ 的光学特性和稳定性。

Characterization of the optical properties and stability of Presage™ following irradiation with photons and carbon ions.

机构信息

Department of Medical Physics, Aarhus University Hospital, Aarhus, Denmark.

出版信息

Acta Oncol. 2011 Aug;50(6):829-34. doi: 10.3109/0284186X.2011.565368.

Abstract

BACKGROUND

The on-going development of both intensity-modulated radiotherapy (IMRT), including the more recent intensity-modulated arc therapy, as well as particle beam therapy, has created a clear need for accurate verification of dose distributions in three dimensions (3D). Presage™ is a new 3D dosimetry material that exhibits a radiochromic response when exposed to ionizing radiation. In this study we have 1) developed an improved optical set-up for measurements of changes in OD of Presage™ point dosimeters, 2) investigated the dose response of Presage™ for photons and carbon ions in the therapy range, 3) investigated the dose response of Presage™ for photons in the kGy range and 4) investigated the fading (i.e. bleaching) of Presage™ postirradiation.

MATERIALS AND METHODS

Presage™ was examined in 1 × 1 × 4.5 cm(3) optical cuvettes; a cuvette holder assured accurate repositioning, and the optical setup included a reference detector to take into account laser intensity fluctuations. The cuvettes were measured pre- and postirradiation for a two week period.

RESULTS

A linear response was observed between dose and optical response between 0 Gy and 100 Gy for γ-radiation from Co-60 and for carbon ions (both plateau and SOBP) from 0 to 20 Gy. The dosimeter was found to have a saturation dose of approximately 100 Gy for photons. A linear energy transfer (LET) effect was not observed in the dose response of different LET radiation. The postirradiation change in optical fading was found to be 0.5% ΔOD/day.

CONCLUSIONS

Our study shows that Presage™ remains a dosimeter of interest for radiation therapy with other particles as well as photons in the therapy dose range.

摘要

背景

调强放射治疗(IMRT)的不断发展,包括最近的强度调制弧形治疗,以及粒子束治疗,都明确需要对三维(3D)剂量分布进行准确验证。Presage™是一种新的 3D 剂量测定材料,在暴露于电离辐射时会表现出放射色响应。在这项研究中,我们 1)开发了一种用于测量 Presage™点剂量计 OD 变化的改进光学设置,2)研究了 Presage™在治疗范围内对光子和碳离子的剂量响应,3)研究了 Presage™在光子的 kGy 范围内的剂量响应,以及 4)研究了 Presage™的后照射褪色(即漂白)。

材料和方法

在 1×1×4.5cm(3)光学比色皿中检查 Presage™;比色皿支架确保了精确的重新定位,光学设置包括一个参考探测器,以考虑激光强度波动。在两周的时间内,对比色皿进行了辐照前后的测量。

结果

在 Co-60 的 γ 射线和 0 到 20Gy 的碳离子(均为平台和 SOBP)的剂量与光学响应之间观察到了从 0Gy 到 100Gy 的线性响应。发现该剂量计对于光子的饱和剂量约为 100Gy。在不同 LET 辐射的剂量响应中未观察到线性能量转移(LET)效应。辐照后光学褪色的变化被发现为 0.5%ΔOD/天。

结论

我们的研究表明,Presage™仍然是一种有前途的治疗剂量范围内其他粒子以及光子的放射治疗剂量计。