Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Kyoto 6068507, Japan.
Med Phys. 2011 May;38(5):2535-41. doi: 10.1118/1.3578607.
To evaluate the accuracy of the patient-positioning function of a newly developed image-guided radiotherapy system, the MHI-TM2000 (Mitsubishi Heavy Industries, Ltd., Japan).
The isocenter positions prescribed by the lasers, MV treatment beam, and image guidance systems (kV X-ray image and kV-CBCT) were calculated using a cube phantom with a 10-mm-diameter steel ball fixed to the center of the phantom. Then, their location discrepancies were estimated. In addition, to verify the scale and orientation of the coordinate axes of the kV X-ray imaging system, positional measurements were repeated with the phantom placed at 50 mm off-isocenter along the vertical, longitudinal, and lateral directions, respectively. Further, image fusions of an anthropomorphic phantom image and the corresponding image translated by a pre-determined amount were performed.
The isocenter alignment among the coordinate systems was coincident within 0.5 mm in translation for the vertical, longitudinal, and lateral axes, respectively. The geometrical errors at 50 mm off-isocenter for kV X-ray images and CBCT were within 0.2 mm and 1.0 mm, respectively. The image fusion errors were within 1.0 mm in translation and 1.0 degrees in rotation, respectively. No significant difference in the image fusion accuracy was observed between the chest and pelvis phantoms.
The isocenter alignment among the coordinate systems was performed with high accuracy. Furthermore, the automatic image fusion function achieved sufficient patient positioning accuracy and precision for image-guided radiotherapy.
评估新开发的图像引导放射治疗系统 MHI-TM2000(日本三菱重工业株式会社)的患者定位功能的准确性。
使用中心固定有 10mm 直径钢球的立方体型模体,计算激光、MV 治疗射线和图像引导系统(kV X 射线图像和 kV-CBCT)规定的等中心点位置,并估计它们的位置差异。此外,为了验证 kV X 射线成像系统的坐标轴的比例和方向,分别将模体放置在离等中心 50mm 的垂直、纵向和横向位置,重复进行位置测量。进一步,对人体模型图像和相应的预先确定量平移的图像进行图像融合。
在垂直、纵向和横向坐标轴上,坐标系的等中心点对齐精度在 0.5mm 以内。kV X 射线图像和 CBCT 在离等中心 50mm 处的几何误差分别在 0.2mm 和 1.0mm 以内。图像融合误差在平移时为 1.0mm,旋转时为 1.0 度,在平移和旋转时分别为 1.0mm。在胸部和骨盆模型中,图像融合精度没有显著差异。
坐标系的等中心点对齐精度很高。此外,自动图像融合功能实现了图像引导放射治疗的足够的患者定位精度和准确性。