Vero4DRT系统的锥形束计算机断层扫描图像质量保证评估

Evaluation of cone-beam computed tomography image quality assurance for Vero4DRT system.

作者信息

Miura Hideharu, Ozawa Shuichi, Hayata Masahiro, Tsuda Shintarou, Enosaki Tsubasa, Yamada Kiyoshi, Nagata Yasushi

机构信息

Hiroshima High-Precision Radiotherapy Cancer Center, Japan.

Department of Radiation Oncology, Institute of Biomedical & Health Science, Hiroshima University, Japan.

出版信息

Rep Pract Oncol Radiother. 2017 May-Jun;22(3):258-263. doi: 10.1016/j.rpor.2016.12.001. Epub 2017 Apr 25.

Abstract

We report the characteristics of quality assurance (QA) image for Vero4DRT system with a kilo-voltage (kV) cone-beam computed tomography (CBCT) capability to perform image-guided radiation therapy (IGRT). To acquire a set of CBCT, the kV source is rotated either 215° clockwise (CW) (tube 1 from 5° to 220° and tube 2 from 275° to 130°) or counterclockwise (CCW) (tube 1 from 85° to 230° and tube 2 from 355° to 140°). Image geometry, image uniformity, high/low contrast resolutions, and contrast linearity were measured with a Catphan 504 CT phantom (The Phantom Laboratory, NY). The comparison between measured and expected distances shows an excellent agreement. The CBCT for Vero4DRT system cannot perform a full 360° rotation, which leads to a loss in uniformity for image acquisition. Separations were observed for high-contrast resolution, with eight line pairs per centimeter corresponding to a gap size of 0.063 cm. For low-contrast resolution, the seventh largest hole was visible. This hole has a 4-mm diameter with 1.0% contrast level. We should check the contrast linearity compared with known value, even though it is out of range from the manufacturer manual.

摘要

我们报告了具有千伏(kV)锥形束计算机断层扫描(CBCT)功能以执行图像引导放射治疗(IGRT)的Vero4DRT系统的质量保证(QA)图像特征。为获取一组CBCT,kV源可顺时针(CW)旋转215°(管1从5°旋转至220°,管2从275°旋转至130°)或逆时针(CCW)旋转(管1从85°旋转至230°,管2从355°旋转至140°)。使用Catphan 504 CT体模(幻影实验室,纽约)测量图像几何形状、图像均匀性、高/低对比度分辨率和对比度线性。测量距离与预期距离之间的比较显示出极佳的一致性。Vero4DRT系统的CBCT无法进行360°全旋转,这导致图像采集的均匀性有所损失。在高对比度分辨率方面观察到有间隔,每厘米八条线对对应于0.063厘米的间隙大小。对于低对比度分辨率,可见第七大的孔。该孔直径为4毫米,对比度水平为1.0%。即使超出制造商手册的范围,我们也应将对比度线性与已知值进行比较。

相似文献

1
Evaluation of cone-beam computed tomography image quality assurance for Vero4DRT system.
Rep Pract Oncol Radiother. 2017 May-Jun;22(3):258-263. doi: 10.1016/j.rpor.2016.12.001. Epub 2017 Apr 25.
2
Image quality and absorbed dose comparison of single- and dual-source cone-beam computed tomography.
J Appl Clin Med Phys. 2018 May;19(3):360-366. doi: 10.1002/acm2.12328. Epub 2018 Apr 17.
3
Technical Note: Assessing the performance of monthly CBCT image quality QA.
Med Phys. 2019 Jun;46(6):2575-2579. doi: 10.1002/mp.13535. Epub 2019 Apr 24.
7
Flat-panel cone-beam computed tomography for image-guided radiation therapy.
Int J Radiat Oncol Biol Phys. 2002 Aug 1;53(5):1337-49. doi: 10.1016/s0360-3016(02)02884-5.
8
Evaluation of imaging performance of major image guidance systems.
Biomed Imaging Interv J. 2011 Apr;7(2):e11. doi: 10.2349/biij.7.2.e11. Epub 2011 Apr 1.

引用本文的文献

1
Comparison of three different phantoms used for Winston-Lutz test with Artiscan software.
Rep Pract Oncol Radiother. 2020 May-Jun;25(3):351-354. doi: 10.1016/j.rpor.2020.03.003. Epub 2020 Mar 4.
2
Image quality and absorbed dose comparison of single- and dual-source cone-beam computed tomography.
J Appl Clin Med Phys. 2018 May;19(3):360-366. doi: 10.1002/acm2.12328. Epub 2018 Apr 17.

本文引用的文献

3
Utilization of cone-beam CT for offline evaluation of target volume coverage during prostate image-guided radiotherapy based on bony anatomy alignment.
Rep Pract Oncol Radiother. 2012 May 5;17(3):134-40. doi: 10.1016/j.rpor.2012.03.003. eCollection 2012.
5
Evaluation of imaging performance of major image guidance systems.
Biomed Imaging Interv J. 2011 Apr;7(2):e11. doi: 10.2349/biij.7.2.e11. Epub 2011 Apr 1.
6
Positioning accuracy of a new image-guided radiotherapy system.
Med Phys. 2011 May;38(5):2535-41. doi: 10.1118/1.3578607.
8
Evaluation of radiation dose and image quality for the Varian cone beam computed tomography system.
Int J Radiat Oncol Biol Phys. 2011 May 1;80(1):291-300. doi: 10.1016/j.ijrobp.2010.06.014.
9
Task Group 142 report: quality assurance of medical accelerators.
Med Phys. 2009 Sep;36(9):4197-212. doi: 10.1118/1.3190392.
10
Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study.
Radiother Oncol. 2009 Oct;93(1):1-7. doi: 10.1016/j.radonc.2009.07.012. Epub 2009 Aug 18.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验