Department of Anaesthesia, University Children's Hospital Zurich, Zurich, Switzerland.
Int J Lab Hematol. 2012 Feb;34(1):86-90. doi: 10.1111/j.1751-553X.2011.01363.x. Epub 2011 Jul 27.
The aim of the study was to compare accuracy and reproducibility of four point-of-care testing (POCT) devices (GEM® Premier 3000, ABL 800 flex, GEM® OPL™, HemoCue® B-Hemoglobin) for hemoglobin (Hb) analyzes as compared with the reference laboratory method (Sysmex XE 2100) in children undergoing major surgery.
Pediatric patients undergoing craniofacial, spine, hip, or cancer surgery were included. Blood samples for Hb testing were taken at several intraoperative time points and generally withdrawn from the arterial catheter, if accessible.
A total of 256 blood samples were taken intraoperatively from 71 pediatric patients. All POCT devices showed very small bias (maximum -0.46 g/dL) to reference method as well as very good reproducibility (maximum coefficient of variation of 0.99%). However, in two cases (HemoCue), potential clinical relevant differences were observed beyond a range of 2 g/dL.
All POCT devices tested and operated by trained staff for hemoglobinometry showed reliable test results. They all allow for simple, fast, and precise bedside determination of hemoglobin concentration in the intraoperative setting.
本研究旨在比较四种即时检测(POCT)设备(GEM® Premier 3000、ABL 800 flex、GEM® OPL™、HemoCue® B-Hemoglobin)与参考实验室方法(Sysmex XE 2100)在接受大型手术的儿童中进行血红蛋白(Hb)分析的准确性和可重复性。
纳入接受颅面、脊柱、髋关节或癌症手术的儿科患者。在几个手术期间的时间点采集用于 Hb 检测的血液样本,通常从动脉导管中抽取,如果可以的话。
共从 71 名儿科患者术中采集了 256 份血液样本。所有 POCT 设备与参考方法相比显示出非常小的偏差(最大 -0.46 g/dL),并且具有非常好的可重复性(最大变异系数为 0.99%)。然而,在两种情况下(HemoCue),观察到潜在临床相关差异超过 2 g/dL 的范围。
所有经过培训的人员操作和测试的 POCT 设备均显示出可靠的测试结果。它们都允许在手术环境中简单、快速和精确地床边确定血红蛋白浓度。