Clin Chim Acta. 2011 Nov 20;412(23-24):2076-84. doi: 10.1016/j.cca.2011.07.009. Epub 2011 Jul 23.
Newborn screening for congenital adrenal hyperplasia (CAH) is commonly accomplished by measurement of 17-α-hydroxyprogestrone (17-OHP) using enzyme immunoassay (EIA). EIA contributes a significant number of false positives. Therefore, second-tier steroid profile by liquid chromatography-tandem mass spectrometry (LC-MS/MS) is warranted.
Dried blood spots (DBS) were extracted with a mixture of methanol and water containing the deuterium labeled internal standards of d(8)-17-OHP, d(7)-androstenedione, and d(4)-cortisol. The final extracts were analyzed for 17-OHP, androstenedione and cortisol by LC-MS/MS in the multiple reaction monitoring (MRM) mode.
Mean recoveries of the target analytes, 17-OHP, androstenedione and cortisol, were between 97 and 115% with an average intra- and inter-assay CVs ranging from 3.9-9.9% to 3.6-10.1%, respectively. The high efficiency of this method enabled us to test 11,598 specimens, identified as indeterminate by EIA in ~6 years; resulting in 809 presumptive positives reducing the false positives rate by 93%.
The three steroid profile provided better screening outcomes of CAH than 17-OHP concentration alone. Our sample preparation allowed high throughput using common laboratory chemicals. Using three internal standards significantly improved method precision and accuracy. The reduction in false positives significantly reduces anxiety for newborns and their families.
先天性肾上腺皮质增生症(CAH)的新生儿筛查通常通过酶免疫分析法(EIA)测量 17-α-羟孕酮(17-OHP)来完成。EIA 会产生大量的假阳性。因此,需要通过液相色谱-串联质谱法(LC-MS/MS)进行二级类固醇谱分析。
用甲醇和水的混合物从干血斑(DBS)中提取,其中含有氘标记的内标 d(8)-17-OHP、d(7)-雄烯二酮和 d(4)-皮质醇。最终提取物采用 LC-MS/MS 在多重反应监测(MRM)模式下分析 17-OHP、雄烯二酮和皮质醇。
目标分析物 17-OHP、雄烯二酮和皮质醇的平均回收率在 97%至 115%之间,平均日内和日间 CVs 分别在 3.9%-9.9%至 3.6%-10.1%之间。该方法效率高,可在 6 年内测试通过 EIA 鉴定为不确定的 11598 个样本,结果有 809 个疑似阳性,使假阳性率降低了 93%。
与仅测量 17-OHP 浓度相比,三种类固醇谱可提供更好的 CAH 筛查结果。我们的样品制备方法使用常见的实验室化学品可实现高通量。使用三个内标可显著提高方法的精密度和准确性。假阳性的减少显著降低了新生儿及其家庭的焦虑。