Department of Pathology, Toronto General Hospital/University Health Network, Toronto, Canada.
Cardiovasc Pathol. 2012 May-Jun;21(3):158-68. doi: 10.1016/j.carpath.2011.05.003. Epub 2011 Aug 3.
The Medtronic Freestyle valve is a stentless porcine valve with reportedly excellent clinical and hemodynamic results, but little has been reported about its long-term pathology.
Seventeen Freestyle valves were explanted (from 2003 to 2009) and reviewed to assess reasons for bioprosthesis failure. All valves were examined in detail, using histochemistry and immunohistochemistry to identify morphological changes, as well as cellular and humoral responses.
One Freestyle valve, explanted for mitral valve endocarditis on the fifth postoperative day, was excluded from analysis. The average implant duration was 71.1±35.2 months. Six valves were explanted for infective endocarditis, six for aortic insufficiency, and four for aortic stenosis. Calcification was seen in 11 explants, pannus in 15, thrombus in 12, cusp tears in 9, and 10 explants showed needle tract-like injuries. A chronic inflammatory reaction involving the xenograft arterial wall was seen in 15 of 16 valves. The cells were composed of macrophages and lymphocytes, including T cells (CD8 positive) and B cells. Significant damage to the porcine aortic wall was seen in 15 cases, and cusp myocardial shelf damage in 7 cases. All cases stained positively for IgG and C4dpar.
The porcine aortic tissue showed T cell-mediated rejection and significant aortic medial damage, consistent with dilatation of the porcine aortic root. The demonstration of IgG suggests the likelihood of humoral rejection, in addition to cellular rejection. One of the underlying possibilities is that the porcine aortic tissues are inadequately fixed, hence the retained antigenicity.
美敦力 Freestyle 瓣膜是一种无支架猪瓣,据报道具有出色的临床和血液动力学效果,但关于其长期病理学的报道甚少。
从 2003 年到 2009 年,共取出 17 个 Freestyle 瓣膜进行分析,以评估生物瓣失败的原因。所有瓣膜均进行了详细检查,使用组织化学和免疫组织化学来识别形态变化,以及细胞和体液反应。
一个 Freestyle 瓣膜因术后第五天二尖瓣心内膜炎而被排除在分析之外。平均植入时间为 71.1±35.2 个月。6 个瓣膜因感染性心内膜炎、6 个因主动脉瓣关闭不全、4 个因主动脉瓣狭窄而被取出。11 个标本有钙化,15 个有瓣叶赘生物,12 个有血栓,9 个有瓣叶撕裂,10 个标本有针道样损伤。16 个瓣膜中有 15 个显示涉及异种移植物动脉壁的慢性炎症反应。这些细胞由巨噬细胞和淋巴细胞组成,包括 T 细胞(CD8 阳性)和 B 细胞。15 例可见明显的猪主动脉壁损伤,7 例可见瓣叶心肌架损伤。所有病例 IgG 和 C4dpar 染色均为阳性。
猪主动脉组织显示 T 细胞介导的排斥反应和明显的主动脉中层损伤,与猪主动脉根部扩张一致。IgG 的存在提示除细胞排斥外,还存在体液排斥的可能性。其中一个潜在的可能性是猪主动脉组织固定不充分,因此保留了抗原性。