A survey of the state of mammography practice in Bulgaria.

作者信息

Avramova-Cholakova S, Vassileva J

机构信息

Radiation Protection at Medical Exposure Section, National Centre of Radiobiology and Radiation Protection, 3 G Sofijski str, Sofia 1606, Bulgaria.

出版信息

Radiat Prot Dosimetry. 2011 Sep;147(1-2):184-6. doi: 10.1093/rpd/ncr320. Epub 2011 Aug 9.

Abstract

The purpose of this work is to compare the state of a representative number of mammography systems in Bulgaria in terms of incident air kerma (IAK) and film optical density (OD) and to propose the optimisation measures where appropriate. IAK with a standard phantom and OD of the exposed films were measured on 22 % of all units in the country. IAK values ranged from 2 to 24.3 mGy, with the median being 7.9 mGy, first and third quartiles respectively being 5 and 11.2 mGy and the mean 8.6 mGy. OD values ranged from 0.5 to 2.4. Only 36 % of all units had OD within the recommended range. Only 27 % were satisfactorily set with OD within the recommended range and IAK below the third quartile. Full quality control (QC) measurements on 12 units showed a lot of problems. The observed big variations of IAK, with а factor of 12, are due to the non-optimised settings and film processing, the lack of standardisation and periodical QC. Corrective actions were recommended to optimise the settings of some systems and to improve the practice.

摘要

这项工作的目的是比较保加利亚一定数量具有代表性的乳腺摄影系统在空气比释动能(IAK)和胶片光学密度(OD)方面的状况,并在适当情况下提出优化措施。在该国所有设备的22%上测量了使用标准体模时的IAK以及所曝光胶片的OD。IAK值范围为2至24.3毫戈瑞,中位数为7.9毫戈瑞,第一和第三四分位数分别为5和11.2毫戈瑞,平均值为8.6毫戈瑞。OD值范围为0.5至2.4。所有设备中只有36%的OD在推荐范围内。只有27%的设备OD设置令人满意,即在推荐范围内且IAK低于第三四分位数。对12台设备进行的全面质量控制(QC)测量显示存在很多问题。观察到IAK有12倍的巨大变化,这是由于设置和胶片处理未优化、缺乏标准化以及定期QC所致。建议采取纠正措施来优化一些系统的设置并改进操作。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验