Guan Zhi-chen, Deng Xiao-lin, Zhang Qian
Department of Urology, Peking University Shenzhen Hospital, Shenzhen 518036, China.
Beijing Da Xue Xue Bao Yi Xue Ban. 2011 Aug 18;43(4):616-9.
To design a new portable home electronic uroflowmeter and compare it with traditional methods.
The system consists of collectors, urine conducting apparatus, intelligent cell phone, wireless network communication technology, computer analysis and drawing, and data storage technology, etc., and can automatically collect voiding information from patients with lower urinary tract symptoms(LUTS) Through Bluetooth, the voiding information was sent to the patient's intelligent cell phone from the collector, then stored directly by intelligent cell phone and wirelessly transmitted to the workstation in hospital. The system was primarily tested with regard to accuracy of measurement of the voided volume. Multiple doses with known volume were introduced in the system and Laborie uroflowmeter. Furthermore, 38 outpatients who had LUTS were tested simultaneously with the system and Laborie uroflowmeter. The statistical method for assessing agreement between the two methods of clinical measurement was Bland-Altman analysis. Among the subjects, there were 22 male patients and 16 female patients, ranging from 21 to 37 years old, with an average age of 25.5 years, of whom, 19 were tested once and 19 patients twice, equaling to 57 tests.
The system could accurately collect and analyze voiding time, uroflowmetry, voided volume, and automatically provide uroflowmetry parameters. The measurement error of 100, 200, 300, 500 and 800 mL is less than 5%. 12.28%, 5.26% and 3.51% of the Qmax, Qave and voided volume points were beyond the 95% limits of agreement. The maximum absolute values of the Qmax, Qave and voided volume difference were 0.38 mL/s, 0.70 mL/s and 2.90 mL, respectively.They agreed with the recommendation of Standardization International Continence Society.
The new portable home electronic uroflowmeter has good agreement with Laborie uroflowmeter,and is a new LUTS monitoring system integrated with correct, reliable, real-time, convenient and easy-managing advantages. It is as noninvasive and reliable as traditional methods, and its portable feature facilitates application out of hospitals. It can also record voiding diaries.
设计一种新型便携式家用电子尿流计,并与传统方法进行比较。
该系统由收集器、导尿装置、智能手机、无线网络通信技术、计算机分析绘图及数据存储技术等组成,可自动收集下尿路症状(LUTS)患者的排尿信息。通过蓝牙将排尿信息从收集器发送至患者的智能手机,然后由智能手机直接存储并无线传输至医院的工作站。该系统主要针对排尿量测量的准确性进行测试。将已知体积的多剂量液体引入该系统和Laborie尿流计中。此外,对38例LUTS门诊患者同时使用该系统和Laborie尿流计进行测试。评估两种临床测量方法一致性的统计方法为Bland-Altman分析。受试者中,男性患者22例,女性患者16例,年龄21至37岁,平均年龄25.5岁,其中19例测试1次,19例测试2次,共计57次测试。
该系统能够准确收集和分析排尿时间、尿流率、排尿量,并自动提供尿流率参数。100、200、300、500和800 mL的测量误差小于5%。Qmax、Qave和排尿量点分别有12.28%、5.26%和3.51%超出95%一致性界限。Qmax、Qave和排尿量差值的最大绝对值分别为0.38 mL/s、0.70 mL/s和2.90 mL。它们符合国际尿控协会标准化的建议。
新型便携式家用电子尿流计与Laborie尿流计具有良好的一致性,是一种集准确、可靠、实时、便捷、易管理等优点于一体的新型LUTS监测系统。它与传统方法一样无创且可靠,其便携性便于在院外应用。它还可以记录排尿日记。