便携式尿流计数据采集与测量准确性的研究

[Research on data acquisition of portable uroflowmeter and accuracy of measurement].

作者信息

Jiang Yaojun, Ge Bin, Yan Rongguo, Ruan Hongyu

出版信息

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2014 Feb;31(1):69-75.

Abstract

This paper presents the preliminary design of data acquisition system of a portable uroflowmeter. The system uses double-hole cantilever pressure sensor. The signal is transferred to ATmega644PA microprogrammed control unit (MCU), converted by A/D (analog to digital) convertor. Then the further data are processed and get the corresponding relationship of weight-time and two curves of urine flow and urinary flow rate. In the measurement accuracy of the device about urine flow, two factors about the placement and height of the data acquisition are analyzed to show the accuracy of the equipment through the Origin 8.0 data analysis software. The design is characterized by low cost and high speed of data collection, real-time, high accuracy.

摘要

本文介绍了一种便携式尿流计数据采集系统的初步设计。该系统采用双孔悬臂式压力传感器。信号被传输到ATmega644PA微程序控制单元(MCU),由A/D(模拟到数字)转换器进行转换。然后对进一步的数据进行处理,得到重量-时间以及尿流量和尿流率两条曲线的对应关系。在该设备关于尿流量的测量精度方面,分析了数据采集的放置和高度这两个因素,通过Origin 8.0数据分析软件展示设备的精度。该设计的特点是成本低、数据采集速度快、实时性强、精度高。