Byrd G D, Fowler K W, Hicks R D, Lovette M E, Borgerding M F
R. J. Reynolds Tobacco Company, Bowman Gray Technical Center, Winston-Salem, NC 27102.
J Chromatogr. 1990 Mar 23;503(2):359-68. doi: 10.1016/s0021-9673(01)81515-6.
A cryogenic trapping method with isotope dilution gas chromatography-mass spectrometry analysis has been developed for the determination of benzene, toluene, styrene and acrylonitrile in mainstream vapor phase cigarette smoke. The method is simple, direct, and quantitative. Vapor phase samples are collected cryogenically in a series of four traps following removal of the particulate phase with a Cambridge filter pad. For all four analytes, 75-85% of the total amounts recovered were found in the initial trap and less than 1% in the final trap. Assessment of instrumental precision by multiple injections of a sample gave relative standard deviations of less than 2%. Linear calibration for all analytes over the analysis range gave an r2 value greater than 0.99 with average relative standard deviations at the mean ranging from 1.4 to 8.2%. The cigarettes analyzed include a reference cigarette (Kentucky 1R4F), a commercial ultra-low "tar" mentholated cigarette, and two cigarettes that heat but do not burn tobacco. The values determined for the four analytes in the 1R4F samples are comparable to reported values of similar cigarettes. The cigarettes which heat rather than burn tobacco yield less of all four analytes compared to the other cigarettes in the study.
已开发出一种结合低温捕集法与同位素稀释气相色谱 - 质谱分析法,用于测定主流气相卷烟烟雾中的苯、甲苯、苯乙烯和丙烯腈。该方法简单、直接且可定量。在用剑桥滤片去除颗粒相后,气相样品在一系列四个捕集阱中进行低温收集。对于所有四种分析物,回收总量的75 - 85%存在于初始捕集阱中,而在最终捕集阱中的含量不到1%。通过多次进样评估仪器精密度,相对标准偏差小于2%。在分析范围内对所有分析物进行线性校准,得到的r2值大于0.99,平均相对标准偏差在均值处为1.4%至8.2%。所分析的卷烟包括一种参比卷烟(肯塔基1R4F)、一种商业超低“焦油”薄荷醇卷烟以及两种加热不燃烧烟草的卷烟。1R4F样品中四种分析物的测定值与类似卷烟的报告值相当。与研究中的其他卷烟相比,加热而非燃烧烟草的卷烟产生的所有四种分析物都较少。