Novo Nordisk A/S, Hillerød, Denmark .
Diabetes Technol Ther. 2011 Dec;13(12):1207-11. doi: 10.1089/dia.2011.0110. Epub 2011 Aug 24.
FlexTouch(®) (Novo Nordisk A/S, Bagsvaerd, Denmark) is a new prefilled insulin pen for people with diabetes, with a novel injection mechanism and no push-button extension at any dose setting. This study compared the injection force of FlexTouch with that of SoloStar(®) (sanofi-aventis, Paris, France) and KwikPen(®) (Eli Lilly & Co., Indianapolis, IN).
Injection force was measured with the manufacturers' recommended needle attached to each pen (NovoFine(®) [Novo Nordisk] 32-gauge tip extra thin wall 6 mm needle for FlexTouch and BD [Franklin Lakes, NJ] MicroFine™ 31-gauge 5 mm needle for SoloStar and KwikPen) during injection of the maximum dose (60 IU for KwikPen and 80 IU for FlexTouch and SoloStar). Injection was performed at three different constant push-button speeds.
FlexTouch had a significantly (P<0.0001) lower injection force than SoloStar and KwikPen at all injection speeds. The mean±SD injection force of FlexTouch was 5.1±0.5 N. At 4, 6, and 8 mm/s push-button speeds, the injection force of SoloStar was 13.5±2.1, 19.1±1.9, and 26.9±2.4 N, respectively, and the injection force of KwikPen was 14.5±1.9, 20.9±1.4, and 28.2±1.4 N, respectively.
The injection mechanism of FlexTouch means that insulin injection is driven by a torque spring and not the thumb pressure of the user. This results in a 62-82% lower injection force with FlexTouch than other prefilled insulin pens.
FlexTouch(®)(诺和诺德公司,丹麦 Bagsvaerd)是一种新型的预填充胰岛素笔,适用于糖尿病患者,具有新型注射机制,在任何剂量设置下都没有按钮延伸。本研究比较了 FlexTouch 与 SoloStar(®)(赛诺菲-安万特公司,法国巴黎)和 KwikPen(®)(礼来公司,印第安纳州印第安纳波利斯)的注射力。
在注射最大剂量(KwikPen 为 60IU,FlexTouch 和 SoloStar 为 80IU)时,使用制造商推荐的针头连接到每支笔(FlexTouch 为 NovoFine(®)[诺和诺德]32 号针头,特细壁 6mm 针头,SoloStar 和 KwikPen 为 BD [新泽西州富兰克林湖]MicroFine™31 号针头,5mm 针头)测量注射力。以三种不同的恒定按钮速度进行注射。
在所有注射速度下,FlexTouch 的注射力均显著(P<0.0001)低于 SoloStar 和 KwikPen。FlexTouch 的平均±SD 注射力为 5.1±0.5N。在 4、6 和 8mm/s 的按钮速度下,SoloStar 的注射力分别为 13.5±2.1、19.1±1.9 和 26.9±2.4N,KwikPen 的注射力分别为 14.5±1.9、20.9±1.4 和 28.2±1.4N。
FlexTouch 的注射机制意味着胰岛素注射是由扭矩弹簧驱动的,而不是用户的拇指压力。这使得 FlexTouch 的注射力比其他预填充胰岛素笔低 62-82%。