Ito Kimiteru, Kurihara Kyoko, Ishibashi Akihiko, Morooka Miyako, Mitsumoto Takuya, Minamimoto Ryogo, Kubota Kazuo
Department of Radiology, National Center of Neurology and Psychiatry, Tokyo, Japan.
Nucl Med Commun. 2011 Oct;32(10):936-40. doi: 10.1097/MNM.0b013e32834a123f.
To determine the cut-off value for distinguishing a normal versus an abnormal right-to-left shunt percentage on lung perfusion scintigraphy using (99m)Tc-macroaggregated albumin (MAA).
Fifty-three patients (eight patients with a right-to-left shunt and 45 without a right-to-left shunt) who underwent MAA whole-body imaging for the evaluation of right-to-left shunts were divided into group 1 (eight patients with brain MAA uptake) and group 2 (45 patients without brain MAA uptake). Moreover, group 2 was subdivided into two categories (groups 2a and 2b) based on the results of lung computed tomography, electrocardiography examinations, and pulmonary function tests. The average and standard deviation (SD) of each group were compared. In addition, we estimated the cut-off value for a normal right-to-left shunt percentage using whole-body imaging.
The average right-to-left shunt percentage values and SD were 23.67±12.17% in group 1, 6.68±1.04% in group 2a, and 6.60±0.84% in group 2b. The shunt percentages of groups 2a and 2b were not significantly different (P=0.77). The estimated normal value (mean±2 SD) of group 2 was 6.64±0.94%. Meanwhile, the cut-off value was estimated as 10% based on the distributions of MAA shunt percentages for groups 1 and 2.
The normal range (mean±2 SD) was 6.64±1.88%. The cut-off value for the normal right-to-left shunt percentage in MAA scintigraphy was 10%.
使用(99m)锝-大颗粒聚合白蛋白(MAA)确定肺灌注闪烁扫描中区分正常与异常右向左分流百分比的临界值。
53例接受MAA全身显像以评估右向左分流的患者(8例有右向左分流,45例无右向左分流)被分为第1组(8例脑内有MAA摄取)和第2组(45例脑内无MAA摄取)。此外,根据肺部计算机断层扫描、心电图检查和肺功能测试结果,将第2组再细分为两类(2a组和2b组)。比较每组的平均值和标准差(SD)。另外,我们利用全身显像估计正常右向左分流百分比的临界值。
第1组右向左分流百分比的平均值和SD为23.67±12.17%,2a组为6.68±1.04%,2b组为6.60±0.84%。2a组和2b组的分流百分比无显著差异(P = 0.77)。第2组的估计正常值(平均值±2SD)为6.64±0.94%。同时,根据第1组和第2组MAA分流百分比的分布,临界值估计为10%。
正常范围(平均值±2SD)为6.64±1.88%。MAA闪烁扫描中正常右向左分流百分比的临界值为10%。