Medical Physics Graduate Program, Duke University, Durham, North Carolina 27710, USA.
Med Phys. 2011 Aug;38(8):4689-99. doi: 10.1118/1.3613153.
Localization prior to delivery of SBRT to free-breathing patients is performed by aligning the planning internal target volume (ITV) from 4DCT with an on-board free-breathing cone-beam CT (FB-CBCT) image. The FB-CBCT image is assumed to also generate an ITV that captures the full range of motion, due to the acquisition spanning multiple respiratory cycles. However, the ITV could potentially be underestimated when the ratio of time spent in inspiration versus time spent in expiration (I/E ratio) deviates from unity. Therefore, the aim of this study was to investigate the effect of variable I/E ratios on the FB ITV generated from a FB-CBCT scan.
This study employed both phantom and patient imaging data. For the phantom study, five periodic respiratory cycles were simulated with different I/E ratios. Six patient respiratory cycles with variable I/E ratios were also selected. All profiles were then programmed into a motion phantom for imaging and modified to exhibit three peak-to-peak motion amplitudes (0.5, 1.0, and 2.0 cm). Each profile was imaged using two spherical targets with 1.0 and 3.0 cm diameters. 2D projections were acquired with full gantry rotation of a kiloVoltage (kV) imager mounted onto the gantry of a modem linear accelerator. CBCT images were reconstructed from 2D projections using a standard filtered back-projection reconstruction algorithm. Quantitative analyses for the phantom study included computing the change in contrast along the direction of target motion as well as determining the area (which is proportional to the target volume) inside of the contour extracted using a Canny edge detector. For the patient study, projection data that were previously acquired under an investigational 4D CBCT slow-gantry imaging protocol were used to generate both FB-CBCT and 4D CBCT images. Volumes were then manually contoured from both datasets (using the same window and level) for quantitative comparison.
The phantom study indicated a reduction in contrast at the inferior edge of the ITV (corresponding to inspiration) as the ratio decreased, for both simulated and patient respiratory cycles. For the simulated phantom respiratory cycles, the contrast reduction of the smallest I/E ratio was 27.6% for the largest target with the smallest amplitude and 89.7% for the smallest target with the largest amplitude. For patient respiratory cycles, these numbers were 22.3% and 94.0%, respectively. The extracted area from inside of the target contours showed a decreasing trend as the I/E ratio decreased. In the patient study, the FB-CBCT ITVs of both lung tumors studied were underestimated when compared with their corresponding 4D CBCT ITV. The underestimations found were 40.1% for the smaller tumor and 24.2% for the larger tumor.
The ITV may be underestimated in a FB-CBCT image when a patient's respiratory pattern is characterized by a disparate length of time spent in inspiration versus expiration. Missing the full target motion information during on-board verification imaging may result in localization errors.
在为自由呼吸患者提供 SBRT 之前,通过将 4DCT 中的计划内靶区(ITV)与机载自由呼吸锥形束 CT(FB-CBCT)图像对齐来实现本地化。由于采集跨越多个呼吸周期,因此假定 FB-CBCT 图像也会生成捕获全范围运动的 ITV。然而,当吸气时间与呼气时间之比(I/E 比)偏离 1 时,ITV 可能会被低估。因此,本研究的目的是研究可变 I/E 比对从 FB-CBCT 扫描生成的 FB ITV 的影响。
本研究同时使用了体模和患者成像数据。对于体模研究,使用具有不同 I/E 比的五个周期性呼吸周期进行模拟。还选择了六个具有可变 I/E 比的患者呼吸周期。然后,将所有轮廓编程到运动体模中进行成像,并修改为表现出三个峰到峰运动幅度(0.5、1.0 和 2.0cm)。使用两个直径为 1.0 和 3.0cm 的球形目标对每个轮廓进行成像。使用安装在现代直线加速器龙门上的千伏(kV)成像仪进行全龙门旋转采集二维投影。使用标准滤波反投影重建算法从二维投影重建 CBCT 图像。体模研究的定量分析包括计算沿目标运动方向的对比度变化,并使用 Canny 边缘检测器确定轮廓内的面积(与目标体积成正比)。对于患者研究,使用先前在研究性 4D CBCT 慢龙门成像方案下采集的投影数据来生成 FB-CBCT 和 4D CBCT 图像。然后从两个数据集(使用相同的窗口和级别)手动描绘体积以进行定量比较。
体模研究表明,随着比值的降低,在 ITV 的下边缘(对应于吸气)处的对比度降低,对于模拟和患者呼吸周期均如此。对于模拟的体模呼吸周期,对于具有最小幅度的最大目标,最小 I/E 比的对比度降低为 27.6%,对于具有最大幅度的最小目标,对比度降低为 89.7%。对于患者呼吸周期,这些数字分别为 22.3%和 94.0%。当 I/E 比降低时,从目标轮廓内部提取的面积呈下降趋势。在患者研究中,与相应的 4D CBCT ITV 相比,研究中的两个肺肿瘤的 FB-CBCT ITV 被低估。较小肿瘤的低估率为 40.1%,较大肿瘤的低估率为 24.2%。
当患者的呼吸模式以吸气和呼气时间不同为特征时,FB-CBCT 图像中的 ITV 可能会被低估。在机载验证成像期间错过完整的目标运动信息可能会导致定位错误。