Faculty of Life Sciences, The University of Manchester, Manchester, UK.
Ophthalmic Physiol Opt. 2011 Nov;31(6):588-94. doi: 10.1111/j.1475-1313.2011.00868.x. Epub 2011 Sep 21.
To explore the repeatability of lower-order and higher-order ocular aberrations measured in patients with keratoconus.
The IRX-3 (Imagine Eyes, Paris, France) aberrometer was used to record lower-order and higher-order aberrations in 31 eyes of 31 patients with keratoconus. Four monocular measurements were taken consecutively for each patient. The aberrometry data were analysed up to the 5th Zernike order for a 4-mm pupil diameter. The data were evaluated using repeated-measures anova and Friedman analyses. Repeatability was analysed using within-subject standard deviation (S(W) ) and the repeatability limit (r) calculated as 1.96 ×√2 × S(w).
Of the 11 aberration terms evaluated, the repeatability of Z (2,0) (mean = 1.36 μm; S(W) = 0.09 μm; r = 0.26 μm); Z (2,±2) RMS (mean = 1.05 μm; S(W) = 0.09 μm; r = 0.24 μm) and Z (4,0) aberrations (mean = 0.34 μm; S(W) = 0.09 μm; r = 0.24 μm) showed the highest variability. In contrast, Z (3,±1) RMS aberrations (mean = 0.85 μm; S(W) = 0.06 μm; r = 0.16 μm) and Z (4,±2) RMS aberrations (mean = 0.40 μm; S(W) = 0.07 μm; r = 0.18 μm) showed comparatively better repeatability.
The lower-order and higher-order aberrations measured in this group of keratoconic patients showed higher levels of variability compared to previous investigations of visually-normal subjects. These results may be of interest to eyecare practitioners involved in the design and fitting of aberration-controlling contact lenses for patients with keratoconus.
探讨圆锥角膜患者低阶和高阶像差的可重复性。
使用 IRX-3(Imagine Eyes,巴黎,法国)像差仪对 31 例 31 只圆锥角膜患者的低阶和高阶像差进行测量。对每位患者连续进行 4 次单眼测量。对 4mm 瞳孔直径的前 5 阶泽尼克像差进行分析。采用重复测量方差分析和 Friedman 分析对像差数据进行评估。使用受试者内标准差(S(W))评估重复性,并将重复性限(r)计算为 1.96×√2×S(w)。
在所评估的 11 项像差中,Z(2,0)(平均值=1.36μm;S(W)=0.09μm;r=0.26μm)、Z(2,±2)RMS(平均值=1.05μm;S(W)=0.09μm;r=0.24μm)和 Z(4,0)像差(平均值=0.34μm;S(W)=0.09μm;r=0.24μm)的可变性最大。相比之下,Z(3,±1)RMS 像差(平均值=0.85μm;S(W)=0.06μm;r=0.16μm)和 Z(4,±2)RMS 像差(平均值=0.40μm;S(W)=0.07μm;r=0.18μm)的重复性更好。
与对视力正常人群的先前研究相比,该组圆锥角膜患者测量的低阶和高阶像差显示出更高的可变性。这些结果可能对视光师在为圆锥角膜患者设计和适配像差矫正隐形眼镜时有所帮助。