Vanbiervliet J, Bellemans J, Verlinden C, Luyckx J P, Labey L, Innocenti B, Vandenneucker H
European Centre for Knee Research, Smith & Nephew, Leuven, Belgium.
J Bone Joint Surg Br. 2011 Oct;93(10):1348-54. doi: 10.1302/0301-620X.93B10.26831.
Complications involving the patellofemoral joint, caused by malrotation of the femoral component during total knee replacement, are an important cause of persistent pain and failure leading to revision surgery. The aim of this study was to determine and quantify the influence of femoral component malrotation on patellofemoral wear, and to determine whether or not there is a difference in the rate of wear of the patellar component when articulated against oxidised zirconium (OxZr) and cobalt-chrome (CoCr) components. An in vitro method was used to simulate patellar maltracking for both materials. Both rates of wear and changes in height on the patellar articular surface were measured. The mean rates of wear measured were very small compared to standard tibiofemoral wear rates. When data for each femoral component material were pooled, the mean rate of wear was 0.19 mm3/Mcycle (sd 0.21) for OxZr and 0.34 mm3/Mcycle (sd 0.335) for CoCr. The largest change in height on each patella varied from -0.05 mm to -0.33 mm over the different configurations. The results suggest that patellar maltracking due to an internally rotated femoral component leads to an increased mean patellar wear. Although not statistically significant, the mean wear production may be lower for OxZr than for CoCr components.
全膝关节置换术中股骨组件旋转不良引起的髌股关节并发症,是导致持续疼痛和手术失败进而需要翻修手术的重要原因。本研究的目的是确定并量化股骨组件旋转不良对髌股关节磨损的影响,并确定髌股组件与氧化锆(OxZr)和钴铬合金(CoCr)组件铰接时,其磨损率是否存在差异。采用体外方法模拟两种材料的髌骨轨迹不良。测量了髌骨磨损率和髌骨关节面高度变化。与标准胫股关节磨损率相比,测得的平均磨损率非常小。当汇总每种股骨组件材料的数据时,OxZr的平均磨损率为0.19mm³/M循环(标准差0.21),CoCr为0.34mm³/M循环(标准差0.335)。在不同构型下,每个髌骨的最大高度变化在-0.05mm至-0.33mm之间。结果表明,股骨组件内旋导致的髌骨轨迹不良会使髌骨平均磨损增加。虽然无统计学意义,但OxZr组件的平均磨损量可能低于CoCr组件。