运动学条件和设计对髌股关节置换假体磨损的影响。

The influence of kinematic conditions and design on the wear of patella-femoral replacements.

作者信息

Maiti Raman, Fisher John, Rowley Liam, Jennings Louise M

机构信息

Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK.

出版信息

Proc Inst Mech Eng H. 2014 Feb;228(2):175-81. doi: 10.1177/0954411913518910. Epub 2014 Jan 29.

Abstract

The success rate of patella-femoral arthroplasty varies between 44% and 90% in 17 years of follow-up. Several studies have been performed previously for assessing the surface wear in the patella-femoral joint. However, they have not included all six degrees of freedom. The aim of this study was to develop a six-axis patella-femoral joint simulator to assess the wear rate for two patellae designs (round and oval dome) at different kinematic conditions. An increase in patellar rotation from 1° to 4° led to a significantly (p<0.049) increased wear rate of round dome from 8.6 mm(3)/million cycles to 12.3 mm(3)/million cycles. The wear rate for oval dome increased from 6.3 mm(3)/million cycles to 14.5 mm(3)/million cycles. However, the increase was nonsignificant (p>0.08). The increase in wear rate was likely due to the higher cross shear. A decrease in patellar medial lateral displacement from passive to constrained resulted in a nonsignificant reduction in wear (p>0.06). There was no significant difference in wear rate between the two patellae designs (p>0.28). The volumetric wear under all conditions was positively correlated with the level of passive patellar tilt (rho>0.8). This is the first report of preclinical wear simulation of patella-femoral joint in a six-axis simulator under different kinematic conditions.

摘要

在17年的随访中,髌股关节置换术的成功率在44%至90%之间。此前已经进行了多项研究来评估髌股关节的表面磨损。然而,这些研究并未涵盖所有六个自由度。本研究的目的是开发一种六轴髌股关节模拟器,以评估两种髌骨设计(圆形和椭圆形穹顶)在不同运动条件下的磨损率。髌骨旋转角度从1°增加到4°导致圆形穹顶的磨损率显著(p<0.049)增加,从8.6立方毫米/百万次循环增加到12.3立方毫米/百万次循环。椭圆形穹顶的磨损率从6.3立方毫米/百万次循环增加到14.5立方毫米/百万次循环。然而,这种增加并不显著(p>0.08)。磨损率的增加可能是由于更高的交叉剪切力。髌骨内外侧位移从被动状态到受限状态的减小导致磨损的非显著降低(p>0.06)。两种髌骨设计之间的磨损率没有显著差异(p>0.28)。在所有条件下,体积磨损与被动髌骨倾斜程度呈正相关(rho>0.8)。这是关于在六轴模拟器中不同运动条件下髌股关节临床前磨损模拟的首份报告。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3500/4361355/2cdb68b59180/10.1177_0954411913518910-fig1.jpg