Daridon J-L, Cassiède M, Paillol J H, Pauly J
Laboratoire des Fluides Complexes, UMR 5150, BP 1155, 64013 Pau Cedex, France.
Rev Sci Instrum. 2011 Sep;82(9):095114. doi: 10.1063/1.3638465.
A quartz crystal viscometer has been developed for measuring viscosity in liquids under pressure. It employs an AT-cut quartz crystal resonator of fundamental frequency 3 MHz inserted in a variable-volume vessel designed for working up to 80 MPa. Viscosity is determined by two methods from resonance frequency and bandwidth measurements along up to eight different overtones. The resonance frequency allows an absolute measurement of the viscosity but leads to an accuracy limited to 5% whereas the bandwidth technique which works in a relative way provides an accuracy of 2%. The techniques were tested by carrying out measurements in two pure compounds: heptane and toluene. Measurement results demonstrate the feasibility of the technique in this viscosity range. The apparatus was also used to determine the viscosity of n-decane with dissolved methane. The results obtained with these mixtures reveal the applicability of the apparatus for reservoir fluids study.
一种用于测量受压液体粘度的石英晶体粘度计已被研制出来。它采用了一个基频为3MHz的AT切割石英晶体谐振器,该谐振器插入一个可变容积的容器中,该容器设计用于高达80MPa的工作压力。通过两种方法根据共振频率和带宽测量来确定粘度,测量范围可达八个不同的泛音。共振频率允许对粘度进行绝对测量,但精度限制在5%,而以相对方式工作的带宽技术提供的精度为2%。通过在两种纯化合物(庚烷和甲苯)中进行测量对这些技术进行了测试。测量结果证明了该技术在这个粘度范围内的可行性。该仪器还用于测定溶解有甲烷的正癸烷的粘度。用这些混合物得到的结果揭示了该仪器在储层流体研究中的适用性。