Ohmori H, Arimoto K, Taki M, Kosaka K, Kumon H, Ohmori H
Shigei Medical Research Hospital, Okayama, Japan.
Nihon Jinzo Gakkai Shi. 1990 Apr;32(4):409-14.
We evaluated a new automated biopsy device for percutaneous renal biopsies under ultrasound guidance, which was recently introduced in Japan for prostate biopsies. This device (Biopty-Gun: Bard Biopty Instrument Uppsala, Sweden) employs a Tru-Cut type smaller needle (18 gauge). We were able to obtain one or two renal tissues in all 57 cases with great ease and in little time. The length of specimen was sufficient (5-17 mm), but the width was thinner than the samples with the Vim-Silverman or Tru-Cut needles. We could achieve a definitive pathological diagnosis in 54 of 57 cases (94.7%), but now, we try to obtain two pieces of tissue for taking more adequate tissue. Only three patients had perirenal hematomas noted by computerized tomography or ultrasonography. We believe that this new automated technique offers a safer and more effective means of obtaining renal tissue.
我们评估了一种新型自动活检装置,用于在超声引导下经皮肾活检,该装置最近在日本被用于前列腺活检。此装置(Biopty-Gun:瑞典乌普萨拉巴德Biopty仪器公司)采用一种Tru-Cut型较细的针(18号)。在所有57例病例中,我们都能轻松且迅速地获取一到两块肾组织。标本长度足够(5 - 17毫米),但宽度比使用Vim-Silverman针或Tru-Cut针获取的样本窄。57例病例中有54例(94.7%)能得出明确的病理诊断,但现在,我们尝试获取两块组织以获取更充足的组织。仅3例患者经计算机断层扫描或超声检查发现肾周血肿。我们认为这种新型自动技术为获取肾组织提供了一种更安全、更有效的方法。