Abstract

OBJECTIVE

Cerebral arteriovenous malformations (AVMs) are vascular lesions that are amenable to various treatment modalities including stereotactic radiosurgery, fractionated radiotherapy, endovascular embolization, microsurgical obliteration or combined modality treatment. A potential complication of endovascular therapy with embolization material is microcatheter entrapment. We report on a patient for whom surgery was combined with endovascular embolization to obliterate an AVM and retrieve an entrapped endovascular microcatheter.

PARTICIPANT

A 52-year-old woman suffered a left parietal hemorrhage from an AVM. She underwent staged endovascular embolization of the lesion using Onyx material. During the second stage of the embolization, the microcatheter (Marathon Flow Directed Microcatheter; eV3 Neurovascular, Inc., Irvine, CA, USA) was retained in the Onyx plug. It was decided to section the microcatheter at the groin and proceed with microsurgical obliteration of the AVM, with removal of the entrapped microcatheter remnant.

INTERVENTION

The AVM was dissected circumferentially allowing the meticulous obliteration of the feeding vessels. A single remaining feeding vessel originating from the distal anterior cerebral artery was identified and suspected to contain the entrapped microcatheter. The location was confirmed using stereotactic guidance (BrainLab, Munich, Germany) and the vessel was then sectioned allowing complete removal of the AVM. The microcatheter (102 cm) was then extracted cranially using gentle traction.

CONCLUSION

This demonstrates the first incidence of microcatheter removal after procedural entrapment in Onyx embolization material.

摘要

目的

脑动静脉畸形(AVM)是一种血管病变,可以采用多种治疗方法,包括立体定向放射外科、分次放射治疗、血管内栓塞、显微手术闭塞或联合治疗。血管内治疗栓塞材料的潜在并发症是微导管嵌顿。我们报告了一例患者,采用手术联合血管内栓塞来闭塞 AVM 并取出嵌顿的血管内微导管。

参与者

一名 52 岁女性因 AVM 出血导致左顶叶出血。她接受了分期血管内栓塞治疗,使用 Onyx 材料。在栓塞的第二阶段,微导管(Marathon Flow Directed Microcatheter; eV3 Neurovascular,Irvine,CA,USA)被保留在 Onyx 栓子中。决定在腹股沟处切断微导管,并进行 AVM 的显微手术闭塞,同时取出嵌顿的微导管残端。

干预措施

AVM 被环形解剖,以便仔细闭塞供血血管。发现并怀疑有一根单独的来自远端大脑前动脉的供血血管含有嵌顿的微导管。使用立体定向引导(BrainLab,慕尼黑,德国)确认了位置,然后切断血管,以便完全切除 AVM。然后将微导管(102cm)轻轻牵拉从颅侧取出。

结论

这是首例在 Onyx 栓塞材料中出现程序性嵌顿后取出微导管的病例。