• 文献检索
  • 文档翻译
  • 深度研究
  • 学术资讯
  • Suppr Zotero 插件Zotero 插件
  • 邀请有礼
  • 套餐&价格
  • 历史记录
应用&插件
Suppr Zotero 插件Zotero 插件浏览器插件Mac 客户端Windows 客户端微信小程序
定价
高级版会员购买积分包购买API积分包
服务
文献检索文档翻译深度研究API 文档MCP 服务
关于我们
关于 Suppr公司介绍联系我们用户协议隐私条款
关注我们

Suppr 超能文献

核心技术专利:CN118964589B侵权必究
粤ICP备2023148730 号-1Suppr @ 2026

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验

多孔金属表面珠子脱落加速膝关节假体失效。

Failure of a knee prosthesis accelerated by shedding of beads from the porous metal surface.

作者信息

Cooke T D, Collins A, Wevers H W

机构信息

Department of Surgery, Queen's University, Kingston, Ontario, Canada.

出版信息

Clin Orthop Relat Res. 1990 Sep(258):204-8.

PMID:2203570
Abstract

A 61-year-old man who received a porous-metal-coated knee implant returned eight months later with chronic synovitis, instability, and loosening of his artificial joint. Subsequently, metal beads were detected in the joint space and soft tissues and were also embedded in the articulating surface of the tibial component. There was scoring of the tibial surface, and polyethylene wear particles were noted in the synovial and fibroconnective-tissue membranes, which had formed beneath the tibial component. Many particles were seen inside giant cells and macrophages. Failure in this case was probably accelerated by the granulomatous response in the soft tissue to wear particles. There was osteolysis rather than new bone growth at the interface with the tibial component. Analysis indicated that poor bead-bonding strength may have initiated the problem. Careful appraisal of the outcome from use of beaded porous-metal-coated devices and assurance of their adequate bonding strength are essential for further progress.

摘要

一名61岁男性接受了多孔金属涂层膝关节植入物,八个月后因慢性滑膜炎、关节不稳定和人工关节松动而复诊。随后,在关节间隙和软组织中检测到金属珠,并且这些金属珠还嵌入了胫骨部件的关节表面。胫骨表面有划痕,在胫骨部件下方形成的滑膜和纤维结缔组织膜中发现了聚乙烯磨损颗粒。在巨细胞和巨噬细胞内可见许多颗粒。该病例的失败可能是由于软组织对磨损颗粒的肉芽肿反应加速所致。在与胫骨部件的界面处出现的是骨质溶解而非新骨生长。分析表明,珠子结合强度不佳可能引发了该问题。仔细评估带珠多孔金属涂层装置的使用效果并确保其具有足够的结合强度对于进一步发展至关重要。

相似文献

1
Failure of a knee prosthesis accelerated by shedding of beads from the porous metal surface.多孔金属表面珠子脱落加速膝关节假体失效。
Clin Orthop Relat Res. 1990 Sep(258):204-8.
2
Role of hydroxyapatite coating in resisting wear particle migration and osteolysis around acetabular components.羟基磷灰石涂层在抵抗髋臼假体周围磨损颗粒迁移和骨溶解中的作用。
Biomaterials. 2005 Jul;26(19):4161-9. doi: 10.1016/j.biomaterials.2004.10.020.
3
Granulomatous reaction and cystic bony destruction associated with high wear rate in a total knee prosthesis.全膝关节假体中与高磨损率相关的肉芽肿反应和囊性骨破坏。
Clin Orthop Relat Res. 1985 Sep(198):224-30.
4
[Case report of extensive metallosis in extra-articular tissues after unicompartmental knee joint replacement].[单髁膝关节置换术后关节外组织广泛金属沉着症的病例报告]
Acta Chir Orthop Traumatol Cech. 2003;70(1):47-50.
5
Etiology of osteolysis around porous-coated cementless total hip arthroplasties.多孔涂层非骨水泥型全髋关节置换术后骨溶解的病因
Clin Orthop Relat Res. 1994 Nov(308):111-26.
6
Implant migration and failure in total knee arthroplasty is related to body mass index and tibial component size.全膝关节置换术中植入物的移位和失败与体重指数及胫骨假体尺寸有关。
J Arthroplasty. 2008 Sep;23(6 Suppl 1):104-9. doi: 10.1016/j.arth.2008.05.020.
7
[The effect of ultra high molecular weight polyethylene particle on the tissues of joint prosthesis].[超高分子量聚乙烯颗粒对关节假体组织的影响]
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2005 Jan;19(1):54-7.
8
[Uncemented knee prosthesis. Results apropos of 58 cases with a minimum of 5-year follow-up].[非骨水泥型膝关节假体。58例至少随访5年的结果]
Rev Chir Orthop Reparatrice Appar Mot. 1995;81(6):505-13.
9
Superior sealing effect of hydroxyapatite in porous-coated implants: experimental studies on the migration of polyethylene particles around stable and unstable implants in dogs.羟基磷灰石在多孔涂层植入物中的优异密封效果:犬稳定和不稳定植入物周围聚乙烯颗粒迁移的实验研究
Acta Orthop. 2005 Jun;76(3):375-85.
10
[The significance of wear and material fatigue in loosening of hip prostheses].[磨损与材料疲劳在髋关节假体松动中的意义]
Orthopade. 1989 Sep;18(5):350-69.

引用本文的文献

1
Bead debonding from a modern cementless total hip femoral component with concomitant taper corrosion.现代非骨水泥型全髋关节股骨部件出现的珠粒脱粘并伴有锥度腐蚀。
Arthroplast Today. 2017 May 6;3(4):215-219. doi: 10.1016/j.artd.2017.04.003. eCollection 2017 Dec.