Abstract

PURPOSE

This study was designed to evaluate inter and intraobserver concordance in gross tumor volume (GTV) delineation on megavoltage CT (MVCT) images in patients with postoperative vault recurrences.

MATERIALS AND METHODS

Three observers delineated GTV on MVCT and CECT and two observers recontoured on MVCT images. Tumor volumes were calculated and correlated using Spearman correlation. The standard deviation of centre of mass was averaged on per patient basis. The ratio of common volume and encompassing volume was used to determine intra and interobserver spatial concordance. Lack of difference in spatial concordance ratio between MVCT and CECT images was used as an index of usability of MVCT images.

RESULTS

Thirty six datasets were available for seven patients. High intraobserver GTV correlation was recorded for observer 1 and 2 (r = 0.93 and r = 0.98; P=0.03 and 0.0001). The average intraobserver spatial concordance ratio was 0.57 and 0.62 respectively. The mean GTV of observers 1, 2 and 3 were 31.6 (18.7-52.2); 28.2 (16.7-51.8) and 46.3 cc (29.1-90.5) respectively. Average standard deviation of centre of mass of all observers was less than 5 mm in either direction. Largest interobserver discordance was observed in anterior, inferior and lateral direction. The interobserver spatial concordance of GTV on MVCT and CECT images was 0.34 and 0.36 (P=0.24) respectively.

CONCLUSION

Moderate to good inter and intraobserver GTV correlation was observed on MVCT images, however, was associated with low interobserver spatial concordance on both MVCT and CECT images. Strategies to improve contouring reproducibility on MVCT and KVCT images are desirable.

摘要

目的

本研究旨在评估术后穹窿复发患者在兆伏级CT(MVCT)图像上大体肿瘤体积(GTV)勾画的观察者间和观察者内一致性。

材料与方法

三名观察者在MVCT和增强CT(CECT)上勾画GTV,两名观察者在MVCT图像上重新勾画。计算肿瘤体积并使用Spearman相关性进行关联。在每位患者的基础上计算质心的标准差平均值。使用共同体积与包含体积的比率来确定观察者内和观察者间的空间一致性。MVCT和CECT图像之间空间一致性比率缺乏差异被用作MVCT图像可用性的指标。

结果

七名患者有36个数据集可用。观察者1和2记录到较高的观察者内GTV相关性(r = 0.93和r = 0.98;P = 0.03和0.0001)。观察者内平均空间一致性比率分别为0.57和0.62。观察者1、2和3的平均GTV分别为31.6(18.7 - 52.2);28.2(16.7 - 51.8)和46.3 cc(29.1 - 90.5)。所有观察者质心的平均标准差在任何方向上均小于5毫米。在前后、下和外侧方向观察到最大的观察者间不一致。MVCT和CECT图像上GTV的观察者间空间一致性分别为0.34和0.36(P = 0.24)。

结论

在MVCT图像上观察到观察者间和观察者内GTV具有中度至良好的相关性,然而,在MVCT和CECT图像上观察者间空间一致性较低。需要采取策略来提高MVCT和千伏级CT(KVCT)图像上轮廓勾画的可重复性。