Regional Maxillofacial Unit, Southern General Hospital, Glasgow, UK.
Int J Oral Maxillofac Surg. 2012 Feb;41(2):155-9. doi: 10.1016/j.ijom.2011.09.020. Epub 2011 Nov 1.
The errors produced by occlusal wafers constructed on casts of the teeth mounted on a standard articulator and an improved orthognathic articulator were investigated by carrying out simulated orthognathic surgery on plastic skulls. The wafers were used to relocate the position of the maxillae of the skulls. The vertical and horizontal displacements of the maxillae were determined from measurements of the positions of markers on the skull and teeth. Comparison of the magnitudes of the actual and intended movements showed that wafers constructed on the standard articulator had systematic prediction errors of up to 5mm, but the improved orthognathic articulator showed much smaller random errors. There was a statistically significant improvement in overall accuracy in predicting maxillary Le Fort I position with the use of the improved orthognathic articulator which the authors recommend for clinical use.
通过在标准牙颌架和改良正颌牙颌架上安装的牙齿模型上制作咬合垫,并在塑料颅骨上进行模拟正颌手术,研究了由其产生的误差。这些咬合垫被用于重新定位颅骨上颌的位置。通过测量颅骨和牙齿上标记物的位置,确定了上颌的垂直和水平位移。实际运动和预期运动幅度的比较表明,在标准牙颌架上制作的咬合垫存在高达 5 毫米的系统预测误差,但改良正颌牙颌架显示出的随机误差要小得多。使用改良正颌牙颌架可以显著提高预测上颌 Le Fort I 位置的整体准确性,作者建议将其用于临床。