BMD measurement and precision: a comparison of GE Lunar Prodigy and iDXA densitometers.

机构信息

University of Wisconsin Osteoporosis Clinical Research Program, Madison, WI, USA.

出版信息

J Clin Densitom. 2012 Jan-Mar;15(1):21-5. doi: 10.1016/j.jocd.2011.08.003. Epub 2011 Nov 9.

Abstract

This study assessed bone mineral density (BMD) comparability and precision using Lunar Prodigy and iDXA densitometers (GE Healthcare, Madison, WI) in adults. Additionally, the utility of supine forearm measurement with iDXA was investigated. Lumbar spine and bilateral proximal femur measurements were obtained in routine clinical manner in 345 volunteers, 202 women and 143 men of mean age 52.5 (range: 20.1-91.6)yr. Seated and supine distal forearm scans were obtained in a subset (n=50). Lumbar spine and proximal femur precision assessments were performed on each instrument following International Society for Clinical Densitometry recommendations in 30 postmenopausal women. BMD at the L1-L4 spine, total proximal femur, and femoral neck was very highly correlated (r(2)≥0.98) between densitometers, as was the one-third radius site (r(2)=0.96). Bland-Altman analyses demonstrated no clinically significant bias at all evaluated sites. BMD precision was similar between instruments at the L1-L4 spine, mean total proximal femur, and femoral neck. Finally, one-third radius BMD measurements in the supine vs seated position on the iDXA were highly correlated (r(2)=0.96). In conclusion, there is excellent BMD correlation between iDXA and Prodigy densitometers. Similarly, BMD precision is comparable with these two instruments.

摘要

本研究评估了 Lunar Prodigy 和 iDXA 骨密度仪(GE Healthcare,麦迪逊,威斯康星州)在成年人中的骨密度(BMD)可比性和精密度。此外,还研究了 iDXA 仰卧前臂测量的实用性。在 345 名志愿者中,202 名女性和 143 名男性,平均年龄为 52.5(范围:20.1-91.6)岁,以常规临床方式获得了腰椎和双侧股骨近端的测量值。在一部分(n=50)中还获得了坐式和仰卧式远端前臂扫描。根据国际临床密度测定学会的建议,对 30 名绝经后妇女的每台仪器进行了腰椎和股骨近端的精密度评估。L1-L4 脊柱、总股骨近端和股骨颈的 BMD 在两种密度仪之间高度相关(r(2)≥0.98),三分之一桡骨部位也高度相关(r(2)=0.96)。Bland-Altman 分析表明,在所有评估部位均无临床显著偏差。在 L1-L4 脊柱、总股骨近端和股骨颈处,仪器之间的 BMD 精密度相似。最后,iDXA 仰卧位与坐位的三分之一桡骨 BMD 测量值高度相关(r(2)=0.96)。总之,iDXA 和 Prodigy 骨密度仪之间存在极好的 BMD 相关性。同样,这两种仪器的 BMD 精密度也相当。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b554/3336200/e88c8e74ab56/nihms-361274-f0001.jpg

相似文献

引用本文的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验