Service de rhumatologie, Inserm U1033, université de Lyon, hôpital Edouard-Herriot, 5, place d'Arsonval, 69437 Lyon, France.
Joint Bone Spine. 2012 Jul;79(4):379-83. doi: 10.1016/j.jbspin.2011.09.012. Epub 2011 Nov 15.
Rheumatoid arthritis is characterized by an early inflammatory related periarticular osteopenia. A new high resolution direct digital X-ray device has been recently developed to provide bone texture analysis which is designed to assess changes in trabecular bone architecture. For the first time, we have evaluated trabecular bone texture impairment in rheumatoid arthritis patients compared to healthy controls.
In this cross-sectional study, the reproducibility was assessed by three separate digital X-rays of the right hand, with repositioning in 14 late rheumatoid arthritis patients and 14 healthy subjects. Then, trabecular bone texture of the MCP2 and MCP3 from patients enrolled in a prospective cohort of 78 rheumatoid arthritis patients was compared with that of 50 healthy subjects, using three texture parameters: Hmean, co-occurrence and run-length.
The coefficients of variation of the high resolution direct digital X-ray measurements ranged from 0.5 to 1.8%. Only the Hmean parameter was significantly decreased in rheumatoid arthritis patients compared to healthy subjects at MCP2 (0.637±0.040 vs. 0.654±0.032, P<0.05) and at MCP3 (0.646±0.044 vs. 0.665±0.037, P<0.05). This reduction was significantly correlated to disease activity.
This study demonstrated both the good reproducibility of the high resolution digital X-ray measurements and the trabecular bone texture impairment at MCP joints in rheumatoid arthritis patients. In addition to provide a high resolution hand radiograph, this technique may represent an interesting tool to easily quantify periarticular osteopenia with a low radiation dose.
类风湿关节炎的特征是早期炎症相关的关节周围骨质疏松症。最近开发了一种新的高分辨率直接数字 X 射线设备,用于提供骨纹理分析,旨在评估小梁骨结构的变化。我们首次评估了类风湿关节炎患者与健康对照组相比,小梁骨纹理受损的情况。
在这项横断面研究中,通过对 14 例晚期类风湿关节炎患者和 14 例健康受试者的右手进行三次独立的数字 X 射线检查,评估了重复性。然后,通过三种纹理参数:Hmean、共现和运行长度,比较了 78 例类风湿关节炎患者前瞻性队列中纳入的患者的 MCP2 和 MCP3 的小梁骨纹理与 50 例健康受试者的 MCP2 和 MCP3 的小梁骨纹理。
高分辨率直接数字 X 射线测量的变异系数范围为 0.5%至 1.8%。仅在 MCP2(0.637±0.040 比 0.654±0.032,P<0.05)和 MCP3(0.646±0.044 比 0.665±0.037,P<0.05)中,类风湿关节炎患者的 Hmean 参数显著低于健康受试者。这种降低与疾病活动度显著相关。
本研究既证明了高分辨率数字 X 射线测量的良好可重复性,也证明了类风湿关节炎患者 MCP 关节的小梁骨纹理受损。除了提供高分辨率的手部射线照片外,该技术还可能是一种很有前途的工具,可通过低辐射剂量轻松定量评估关节周围骨质疏松症。