Br J Radiol. 2012 Jul;85(1015):910-6. doi: 10.1259/bjr/74825285. Epub 2011 Nov 17.
To compare reader ratings of the clinical diagnostic quality of 50 and 100 μm computed radiography (CR) systems with screen-film mammography (SFM) in operative specimens.
Mammograms of 57 fresh operative breast specimens were analysed by 10 readers. Exposures were made with identical position and compression with three mammographic systems (Fuji 100CR, 50CR and SFM). Images were anonymised and readers blinded to the CR system used. A five-point comparative scoring system (-2 to +2) was used to assess seven quality criteria and overall diagnostic value. Statistical analysis was subsequently performed of reader ratings (n = 16,925).
For most quality criteria, both CR systems were rated as equivalent to or better than SFM. The CR systems were significantly better at demonstrating skin edge and background tissue (p < 1 × 10(-5)). Microcalcification was best demonstrated on the CR50 system (p < 1 × 10(-5)). The overall diagnostic value of both CR systems was rated as being as good as or better than SFM (p < 1 × 10(-5)).
In this clinical setting, the overall diagnostic performance of both CR systems was as good as or better than SFM, with the CR50 system performing better than the CR100.
比较 50μm 和 100μm 计算机放射摄影(CR)系统与屏片式乳腺摄影(SFM)在手术标本中临床诊断质量的读者评分。
对 57 个新鲜手术乳腺标本的乳腺 X 线照片进行了 10 位读者的分析。使用三种乳腺摄影系统(富士 100CR、50CR 和 SFM)进行相同位置和压缩的曝光。图像经过匿名处理,读者对使用的 CR 系统一无所知。使用五点比较评分系统(-2 至+2)来评估七个质量标准和整体诊断价值。随后对读者评分(n=16925)进行了统计分析。
对于大多数质量标准,两种 CR 系统的评分均等同于或优于 SFM。CR 系统在显示皮肤边缘和背景组织方面明显更好(p<1×10(-5))。微钙化在 CR50 系统上显示最佳(p<1×10(-5))。两种 CR 系统的整体诊断价值均被评为等同于或优于 SFM(p<1×10(-5))。
在这种临床环境下,两种 CR 系统的整体诊断性能均等同于或优于 SFM,而 CR50 系统的性能优于 CR100 系统。