Yokoi Tomohiro, Takata Tadanori, Ichikawa Katsuhiro
Department of Radiology, Kanazawa University Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi. 2011;67(11):1415-25. doi: 10.6009/jjrt.67.1415.
The purpose of this study was to investigate image quality identification methods among direct- and indirect-type flat panel detectors (FPDs) and a computed radiography (CR) system using two radiation qualities RQA3 and RQA5 defined in the IEC 61267 standard. For each system, the digital characteristic curve, the presampled modulation transfer function (MTF), and the normalized noise power spectrum (NNPS) were measured. Images for a burger phantom and a foot-bone phantom were processed by resolution identification utilizing two-dimensional Fourier transform, and then contrast-to-noise ratio (CNR) for each image was measured. For the RQA3, the direct FPD system indicated the highest DQE value, and for the RQA5 DQE value of indirect FPD system, it was a little higher than that of direct FPD system. The CNR results with the resolution identification displayed good accordance with the DQE results in both phantoms. From the DQE results, dose ratios for image quality identification were determined, and the CNRs of the dose-adjusted images were measured. The results for, the CNRs of all systems showed good coincidence. From these findings, they indicated that the DQE measurement is effective to determine the exposure parameters for equalizing the image quality of different types of radiographic systems.
本研究的目的是使用IEC 61267标准中定义的两种辐射质量RQA3和RQA5,研究直接型和间接型平板探测器(FPD)以及计算机X线摄影(CR)系统之间的图像质量识别方法。对于每个系统,测量了数字特性曲线、预采样调制传递函数(MTF)和归一化噪声功率谱(NNPS)。通过利用二维傅里叶变换的分辨率识别对汉堡模型和足部骨骼模型的图像进行处理,然后测量每个图像的对比度噪声比(CNR)。对于RQA3,直接FPD系统显示出最高的DQE值,对于RQA5,间接FPD系统的DQE值略高于直接FPD系统。分辨率识别的CNR结果与两个模型中的DQE结果显示出良好的一致性。根据DQE结果确定图像质量识别的剂量比,并测量剂量调整后图像的CNR。所有系统的CNR结果显示出良好的一致性。从这些发现表明,DQE测量对于确定用于均衡不同类型射线照相系统图像质量的曝光参数是有效的。