Department of Emergency Medicine, The Ohio State University College of Medicine, Columbus, OH 43210, USA.
J Ultrasound Med. 2011 Dec;30(12):1649-55. doi: 10.7863/jum.2011.30.12.1649.
Ultrasound image interpretation and education relies on obtaining a high-quality ultrasound image; however, no literature exists to date attempting to define a high-quality ultrasound image. The purpose of this study was to design and perform a pilot reliability study of the Brightness Mode Quality Ultrasound Imaging Examination Technique (B-QUIET) method for ultrasound quality image assessment.
A single sonologist performed a Trinity hypotensive ultrasound protocol on 3 participants of varying body types. Each participant's ultrasound examination was repeated in 4 locations; static clinic location, mobile ambulance, airplane, and helicopter. Images were reviewed by a sonographer, radiologist, and emergency medicine physician using the B-QUIET method and underwent statistical analysis using generalizability theory for reliability of the assessments using the tool.
The B-QUIET method showed high reliability of most subscale items. Approximately two-thirds of the reviewed images had complete inter-rater reliability on 90% of the items. There was relatively low inter-rater reliability for the Identification/ Orientation subscale items. The inter-rater reliability κ value was calculated as 0.676 overall for the method.
The need for a standardized method to evaluate the quality of an ultrasound image is well documented. The B-QUIET method represents the first attempt to quantify the sonographer component of ultrasound images. Further reliability and validation studies of this method will be needed; however, it represents a tool for standardized ultrasound interpretation, ultrasound training, and institutional quality assessment.
超声图像的解读和教学依赖于获取高质量的超声图像;然而,目前尚无文献试图定义高质量的超声图像。本研究的目的是设计和进行一项关于 Brightness Mode Quality Ultrasound Imaging Examination Technique(B-QUIET)方法用于超声质量图像评估的可靠性初步研究。
一位超声科医生对 3 名不同体型的参与者进行了 Trinity 降压超声检查。每位参与者的超声检查在 4 个位置重复进行;分别是常规诊所位置、移动救护车、飞机和直升机。图像由超声科医生、放射科医生和急诊医师使用 B-QUIET 方法进行评估,并使用概化理论进行统计分析,评估工具的评估可靠性。
B-QUIET 方法显示了大多数子量表项目的高度可靠性。大约三分之二的被评估图像在 90%的项目上具有完整的评分者间可靠性。识别/定位子量表项目的评分者间可靠性相对较低。该方法的评分者间可靠性 κ 值总体计算为 0.676。
需要一种标准化的方法来评估超声图像的质量,这一点已得到充分证明。B-QUIET 方法代表了量化超声图像中超声科医生部分的首次尝试。需要进一步对该方法进行可靠性和验证研究;然而,它代表了标准化超声解读、超声培训和机构质量评估的工具。