European Multicentre Evaluation of the ESAT 6660.

作者信息

Römer M, Haeckel R, Bonini P, Ceriotti G, Vassault A, Solere P, Morer P

机构信息

Zentralkrankenhaus St.-Jürgen-Strasse, Bremen, Germany.

出版信息

J Clin Chem Clin Biochem. 1990 Jun;28(6):435-43.

PMID:2212961
Abstract

The analytical performance of the glucose analyser ESAT 6660 from Eppendorf was studied according to the ECCLS guidelines and partly according the SFBC protocol in a multicentre evaluation involving laboratories from three European countries. The glucose determination in serum and in haemolysate was studied. The following results were obtained. 1. The precision was as good as or better than the precision of the comparison instruments. The coefficients of variation were between 1.1 and 3.4% for the between-days imprecision and between 0.35 and 1.45% for the within-run imprecision experiment. 2. The recovery of control sera values compared with the hexokinase method was between 94.3 and 102.6%. 3. With patient specimens as good agreement was found between the results obtained with the ESAT 6660 and the different comparison instruments (ASTRA, Hitachi 737 and ACP 5040). 4. A drift effect of 1.1-2.3% occurred in 5 of 21 experiments, depending on the individual enzyme membrane. 5. Sample carry-over was not observed. 6. A linearity between 0.5 and 50 mmol/l was found, exceeding the manufacturer's claims. 7. Several different endogenous and exogenous interferences were investigated. No interfering effect was detected for endogenous substances. A positive interference was observed by ascorbic acid at a concentration above 350 mg/l. 8. The practicability of the instrument was judged as very good. It was considered as a disadvantage that the instrument is not capable of piercing sample lids. Also the numeration of samples is not very convenient.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验