Department of Physics, San Diego State University, San Diego, California 92182, USA.
Med Phys. 2012 Jan;39(1):492-502. doi: 10.1118/1.3671941.
To assess the temporal and spatial accuracy of the GateCT™ system (VisionRT, London, UK), a recently released respiratory tracking system for 4DCT, under both ideal and nonideal respiratory conditions.
Three experiments were performed by benchmarking and comparing its results with the ground-truth input data and those generated by the widely used Varian RPM™ system (Real-time Position Management, Varian, Palo Alto, CA). The first experiment used 10 sinusoidal breathing patterns (constant amplitude and frequency using sin(6)ωt), 10 "consistent" patient breathing patterns, and 10 "sporadic" patient breathing patterns. Motion was simulated with the quasar™ Programmable Respiratory Motion Platform (MODUS, London, Canada) as the surrogate. The GateCT™ and RPM™ systems were used to track the breathing patterns. The data from both systems were then analyzed in the Fourier domain, to evaluate temporal/phase accuracy, using the Pearson's correlation coefficient (PCC). The analysis correlated the ground-truth input data against the GateCT™ and RPM™ tracking results, respectively. The second experiment used 10 ideal sinusoidal breathing patterns, five of period 2.0 s, and five of period 5.0 s, with varying abdominal amplitudes found in clinical cases (peak-to-peak range: 1.67-10 mm) to test the sensitivity of the system to reconstruct various range of motion. And, the third experiment used 12 consecutive clinical patients to track the abdominal motion simultaneously by the GateCT™ and RPM™ systems. The baseline of the tracking results from both the two systems was analyzed via the mean-position-estimate (MPE) calculations. All experiments were tracked for at least 120 s.
In the first experiment, the average PCC values (±SD) of all thirty breathing patterns were 0.9995 ± 0.00035 and 0.9994 ± 0.00041 for the GateCT™ and the RPM™ system, respectively. These nearly identical results demonstrated similar temporal/phase tracking accuracy for the two systems. The results in the second experiment, however, revealed a pattern for the GateCT™ system in which the uncertainty of its mean-position tracking increased as the amplitude of the breathing pattern decreased. For example, a non-negligible baseline drift of up to 29.3% with respect to the peak-to-peak amplitude of 1.67-mm was observed. On the contrary, the RPM™ system displayed a more consistent recording of amplitudes over time with the greatest drift being <7.7%. The third experiment confirmed these findings in the clinical setting. Consistent decrease in PCC values due to the increase in artificial amplitude drifts, as the breathing amplitude decreased, was found. The lowest PCC value was 0.7239 for a patient with 1.57-mm peak-to-peak amplitude.
The GateCT™ system revealed its consistency in temporal/phase tracking but had limitations in accurately tracking the absolute abdominal positions, thus suggesting its appropriateness for phase-sorting of 4DCT rather than amplitude-sorting. In contrast, the RPM™ system demonstrated stable respiratory signal tracking in all ranges and accurately both in phase and amplitude, and is a robust system to use for both phase-sorting and amplitude-sorting techniques. The impact of the observed mean-position drift in the GateCT™ system on the resulting 4DCT image quality, in amplitude-sorting, needs further investigation.
评估最近发布的用于 4DCT 的呼吸跟踪系统 GateCT™ 系统(VisionRT,伦敦,英国)在理想和非理想呼吸条件下的时空准确性。
通过基准测试并将其结果与地面实况输入数据和广泛使用的 Varian RPM™ 系统(实时位置管理,Varian,Palo Alto,CA)生成的结果进行比较,进行了三个实验。第一个实验使用了 10 个正弦呼吸模式(使用 sin(6)ωt 恒定幅度和频率)、10 个“一致”患者呼吸模式和 10 个“随机”患者呼吸模式。运动是使用 quasar™可编程呼吸运动平台(MODUS,伦敦,加拿大)作为替代物模拟的。GateCT™和 RPM™系统用于跟踪呼吸模式。然后分别在傅里叶域中分析来自这两个系统的数据,以使用 Pearson 相关系数(PCC)评估时间/相位准确性。分析将地面实况输入数据与 GateCT™和 RPM™跟踪结果分别进行相关。第二个实验使用了 10 个理想正弦呼吸模式,其中 5 个周期为 2.0 s,5 个周期为 5.0 s,具有在临床病例中发现的不同腹部幅度(峰峰值范围:1.67-10 mm),以测试系统对各种运动范围的灵敏度。并且,第三个实验使用 12 个连续的临床患者同时通过 GateCT™和 RPM™系统跟踪腹部运动。两个系统的跟踪结果的基线通过平均位置估计(MPE)计算进行分析。所有实验均至少跟踪 120 s。
在第一个实验中,所有三十个呼吸模式的平均 PCC 值(±SD)分别为 0.9995 ± 0.00035 和 0.9994 ± 0.00041,对于 GateCT™和 RPM™系统,分别。这些几乎相同的结果表明两个系统具有相似的时间/相位跟踪精度。然而,第二个实验的结果显示了 GateCT™系统的一种模式,其中随着呼吸模式幅度的减小,其平均位置跟踪的不确定性增加。例如,观察到高达 29.3%的与 1.67-mm峰峰值幅度相关的不可忽略的基线漂移。相反,RPM™系统显示出随着时间的推移更一致的幅度记录,最大漂移小于 7.7%。第三个实验在临床环境中证实了这些发现。随着呼吸幅度的减小,由于人工幅度漂移的增加,发现 PCC 值一致降低。对于峰峰值幅度为 1.57-mm 的患者,最低 PCC 值为 0.7239。
GateCT™系统显示出其在时间/相位跟踪方面的一致性,但在准确跟踪绝对腹部位置方面存在局限性,因此表明其适用于 4DCT 的相位排序而不是幅度排序。相比之下,RPM™系统在所有范围内均表现出稳定的呼吸信号跟踪,并且在相位和幅度上都非常准确,是一种用于相位排序和幅度排序技术的强大系统。在幅度排序中,需要进一步研究在 GateCT™系统中观察到的平均位置漂移对最终 4DCT 图像质量的影响。