Belanger Maxwell, Saleh Ziad, Volpe Tom, Margiasso Rich, Li Xiang, Chan Maria, Zhu Xiaofeng, Tang Xiaoli
Columbia University; University of Wisconsin Madison.
J Appl Clin Med Phys. 2016 Jul 8;17(4):223-234. doi: 10.1120/jacmp.v17i4.6152.
Calypso L-shaped Surface Beacon transponder has recently become available for clinical applications. We herein conduct studies to validate the Surface Beacon transponder in terms of stability, reproducibility, orientation sensitivity, cycle rate dependence, and respiratory waveform tracking accuracy. The Surface Beacon was placed on a Quasar respiratory phantom and positioned at the isocenter with its two arms aligned with the lasers. Breathing waveforms were simulated, and the motion of the transponder was tracked. Stability and drift analysis: sinusoidal waveforms (200 cycles) were produced, and the amplitudes of phases 0% (inhale) and 50% (exhale) were recorded at each breathing cycle. The mean and standard deviation (SD) of the amplitudes were calculated. Linear least-squares fitting was performed to access the possible amplitude drift over the breathing cycles. Reproducibility: similar setting to stability and drift analysis, and the phantom generated 100 cycles of the sinusoidal waveform per run. The Calypso system's was re-setup for each run. Recorded amplitude and SD of 0% and 50% phase were compared between runs to assess contribution of Calypso electromagnetic array setup variation. Beacon orientation sensitivity: the Calypso tracks sinusoidal phantom motion with a defined angular offset of the beacon to assess its effect on SD and peak-to-peak amplitude. Rate dependence: sinusoidal motion was generated at cycle rates of 1 Hz, .33 Hz, and .2 Hz. Peak-to-peak displacement and SDs were assessed. Respiratory waveform tracking accuracy: the phantom reproduced recorded breathing cycles (by volunteers and patients) were tracked by the Calypso system. Deviation in tracking position from produced waveform was used to calculate SD throughout entire breathing cycle. Stability and drift analysis: Mean amplitude ± SD of phase 0% or 50% were 20.01 ± 0.04 mm and -19.65 ± 0.08 mm, respectively. No clinically significant drift was detected with drift measured as 5.1 × 10-5 mm/s at phase 0% and -6.0 × 10-5 mm/s at phase 50%. Reproducibility: The SD of the setup was 0.06 mm and 0.02 mm for phases 0% and 50%, respectively. The combined SDs, including both setup and intrarun error of all runs at phases 0% and 50%, were 0.07mm and 0.11 mm, respectively. Beacon orientation: SD ranged from 0.032mm to 0.039 mm at phase 0% and from 0.084 mm to 0.096 mm at phase 50%. The SD was found not to vary linearly with Beacon angle in the range of 0° and 15°. A positive systematic error was observed with amplitude 0.07 mm/degree at phase 0% and 0.05 mm/degree at phase 50%. Rate dependence: SD and displacement amplitudes did not vary significantly between 0.2 Hz and 0.33 Hz. At 1 Hz, both 0% and 50% amplitude measurements shifted up appreciably, by 0.72 mm and 0.78mm, respectively. As compared with the 0.33 Hz data, SD at phase 0% was 1.6 times higher and 5.4 times higher at phase 50%. Respiratory waveform tracking accuracy: SD of 0.233 mm with approximately normal distribution in over 134 min of tracking (201468 data points). The Surface Beacon transponder appears to be stable, accurate, and reproducible. Submillimeter resolution is achieved throughout breathing and sinusoidal waveforms.
Calypso L形体表信标应答器最近已可用于临床应用。我们在此进行研究,以验证体表信标应答器在稳定性、可重复性、方向敏感性、周期率依赖性和呼吸波形跟踪准确性方面的性能。将体表信标放置在Quasar呼吸模体上,并将其置于等中心位置,使其双臂与激光对齐。模拟呼吸波形,并跟踪应答器的运动。稳定性和漂移分析:生成正弦波形(200个周期),并在每个呼吸周期记录0%(吸气)和50%(呼气)相位的振幅。计算振幅的平均值和标准差(SD)。进行线性最小二乘拟合,以确定呼吸周期内可能的振幅漂移。可重复性:设置与稳定性和漂移分析类似,模体每次运行生成100个正弦波形周期。每次运行时重新设置Calypso系统。比较各次运行之间记录的0%和50%相位的振幅和SD,以评估Calypso电磁阵列设置变化的影响。信标方向敏感性:Calypso以信标定义的角度偏移跟踪正弦模体运动,以评估其对SD和峰峰值振幅的影响。频率依赖性:以1 Hz、0.33 Hz和0.2 Hz的周期率生成正弦运动。评估峰峰值位移和SD。呼吸波形跟踪准确性:Calypso系统跟踪模体再现的记录呼吸周期(由志愿者和患者提供)。计算整个呼吸周期内跟踪位置与生成波形的偏差的SD。稳定性和漂移分析:0%或50%相位的平均振幅±SD分别为20.01±0.04 mm和-19.65±0.08 mm。在0%相位时漂移测量值为5.1×10-5 mm/s,在50%相位时为-6.0×10-5 mm/s,未检测到具有临床意义的漂移。可重复性:0%和50%相位设置的SD分别为0.06 mm和0.02 mm。包括所有运行在0%和50%相位的设置和运行内误差的组合SD分别为0.07 mm和0.11 mm。信标方向:0%相位时SD范围为0.032 mm至0.039 mm,50%相位时为0.084 mm至0.096 mm。在0°至15°范围内,发现SD与信标角度不成线性变化。在0%相位时观察到正系统误差,振幅为0.07 mm/度,在50%相位时为0.05 mm/度。频率依赖性:在0.2 Hz和0.33 Hz之间,SD和位移振幅没有显著变化。在1 Hz时,0%和50%振幅测量值均明显上移,分别为0.72 mm和0.78 mm。与0.33 Hz数据相比,0%相位的SD高1.6倍,50%相位高5.4倍。呼吸波形跟踪准确性:在超过134分钟的跟踪(201468个数据点)中,SD为0.233 mm,分布近似正态。体表信标应答器似乎稳定、准确且可重复。在整个呼吸和正弦波形中实现了亚毫米分辨率。