Hemodynamic comparison of Hancock and Carpentier-Edwards mitral bioprosthetic valves.

作者信息

Khan S, Mitchell R S, Derby G C, Oyer P E, Miller D C

机构信息

Division of Cardiology, Cedars-Sinai Medical Center, UCLA School of Medicine 90048.

出版信息

Circulation. 1990 Nov;82(5 Suppl):IV75-81.

PMID:2225438
Abstract

Although it is known that pressure gradients and calculated valve areas in bioprosthetic valves are highly flow dependent, no studies have compared bioprosthetic valve performances while adjusting for differences in flow rate. We therefore studied 75 patients undergoing mitral valve replacement who were randomized to receive either Hancock (n = 35) or Carpentier-Edwards (n = 40) bioprosthetic valves. Pressure gradients were measured using transducer-tipped catheters to record left ventricular and left atrial pressures and cardiac outputs by thermodilution. Repeated measurements were made in each patient after either pacing, fluid infusion, or pharmacological intervention to vary flow rates for a total of 239 measurements (mean, 3.2 measurements per patient). Using analyses of variance and covariance, mean valve gradients and the calculated Gorlin area were adjusted for flow rate, valve size, valve type, and interpatient differences to compare hemodynamics. Without flow and interpatient adjustment, the univariate analysis suggested higher mean gradients in the Carpentier-Edwards 29-mm valves (p = 0.038), with a trend toward higher gradients and smaller areas in the Hancock 33-mm valves (p = 0.057 and 0.059, respectively). After adjustment for flow rate and interpatient differences, however, there was no difference at any valve size in the mean pressure gradients (p = 0.13-0.89) or Gorlin valve areas (p = 0.34-0.96). Although measurements within a given patient were consistent, marked interpatient variabilities in gradients and areas were observed for identical valve types and sizes, which were as significant as flow-dependent or size-dependent changes. We conclude that comparisons of valve performance should adjust for variations in flow rate and for interpatient differences with the use of repeated-measures designs.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验