Suppr超能文献

O型臂注册手术导航仪的技术准确性

Technical accuracy of an O-arm registered surgical navigator.

作者信息

Koivukangas Tapani, Katisko Jani P A, Koivukangas John P

机构信息

Department of Mechanical Engineering, University of Oulu, Oulu, Finland.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2011;2011:2148-51. doi: 10.1109/IEMBS.2011.6090402.

Abstract

The objective of this study was to assess the accuracy of a commercial surgical navigator using optical tracking modality with automated registration between O-arm images and the scanned object. Automated registration was enabled by using the spine navigation software of the navigator. The used phantom was designed by the authors of this paper. The surgical navigators and the O-arm are routinely used at Oulu University Hospital, Oulu, Finland. The distances measured with the surgical navigator from the fixed origin of the phantom were compared to the known phantom accuracy assessment coordinates. The error of the surgical navigator was the difference between measured and true values. The mean displacement error was 0.20 mm with a standard deviation of 0.14 mm. The results show that automated registration is very reliable for image guided surgery (IGS) and that the present accuracy assessment method can be used to periodically check surgical navigator accuracy using O-arm data.

摘要

本研究的目的是评估一种商业手术导航仪的准确性,该导航仪采用光学跟踪模式,能在O型臂图像与扫描对象之间进行自动配准。通过使用导航仪的脊柱导航软件实现自动配准。所使用的模型由本文作者设计。芬兰奥卢大学医院常规使用手术导航仪和O型臂。将手术导航仪从模型固定原点测量的距离与已知的模型精度评估坐标进行比较。手术导航仪的误差为测量值与真实值之间的差值。平均位移误差为0.20毫米,标准差为0.14毫米。结果表明,自动配准对于图像引导手术(IGS)非常可靠,并且目前的精度评估方法可用于定期使用O型臂数据检查手术导航仪的精度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验