Lee Hyo-Ki, Kim Joo-Han, Myoung Hyoun-Seok, Lee Jung-Hun, Lee Kyoung-Joung
Department of Biomedical Engineering, Yonsei University, Republic of Korea.
Annu Int Conf IEEE Eng Med Biol Soc. 2011;2011:5195-8. doi: 10.1109/IEMBS.2011.6091285.
This study is to evaluate the repeatability of the accelerometric-method to detect step events for hemiparetic stroke patients. To evaluate this method, four adults with chronic hemiparetic stroke were participated. The repeatability of this method using a single three-axis accelerometer was evaluated with a six optical camera motion capture system. The correlation statistics and Bland-Altman plot were then used to evaluate the agreement between the step-time differences from the accelerometer data and the reflective markers data. The correlation coefficient of each two data was 0.99 (p < 0.001) and retest result was 0.99 (p < 0.001). The mean ± standard deviation (SD) between each two data along with the 95% limits of agreement (LOA = ± 1.96 SD) was 2.58 ± 2.37 ms (LOA = -2.07 ms and 7.23 ms), and retest result was 3.73 ± 2.02 ms (LOA = -0.22 ms and 7.68 ms). These results show that the suggested method is useful to detect step events for hemiparetic stroke patients.