Jules-Gonin Eye Hospital, Lausanne, Switzerland.
Invest Ophthalmol Vis Sci. 2012 Mar 9;53(3):1219-24. doi: 10.1167/iovs.11-8659.
To investigate the effect of intraocular straylight (IOS) induced by white opacity filters (WOF) on threshold measurements for stimuli employed in three perimeters: standard automated perimetry (SAP), pulsar perimetry (PP) and the Moorfields motion displacement test (MDT).
Four healthy young (24-28 years old) observers were tested six times with each perimeter, each time with one of five different WOFs and once without, inducing various levels of IOS (from 10% to 200%). An increase in IOS was measured with a straylight meter. The change in sensitivity from baseline was normalized, allowing comparison of standardized (z) scores (change divided by the SD of normative values) for each instrument.
SAP and PP thresholds were significantly affected (P < 0.001) by moderate to large increases in IOS (50%-200%). The drop in motion displacement (MD) from baseline with WOF 5, was approximately 5 dB, in both SAP and PP which represents a clinically significant loss; in contrast the change in MD with MDT was on average 1 minute of arc, which is not likely to indicate a clinically significant loss.
The Moorfields MDT is more robust to the effects of additional straylight in comparison with SAP or PP.
研究白浊滤光片(WOF)引起的眼内杂散光(IOS)对三种视野计(标准自动视野计、脉冲视野计和 Moorfields 运动位移试验)刺激阈值测量的影响。
4 名健康年轻(24-28 岁)观察者使用每种视野计测试 6 次,每次使用 5 种不同 WOF 中的一种,且不使用 WOF 各 1 次,从而诱导不同程度的 IOS(10%-200%)。使用杂散光仪测量 IOS 的增加。将灵敏度从基线的变化归一化,允许对每个仪器的标准化(z)分数(变化除以规范值的 SD)进行比较。
SAP 和 PP 阈值受到中度至高度 IOS 增加(50%-200%)的显著影响(P < 0.001)。在 SAP 和 PP 中,WOF5 引起的运动位移(MD)从基线下降约 5dB,这代表了临床显著的损失;相比之下,MDT 的 MD 变化平均为 1 分钟弧,这不太可能表示临床显著的损失。
与 SAP 或 PP 相比,Moorfields MDT 对额外杂散光的影响更具鲁棒性。