Sci Rep. 2017 Jun 12;7(1):3293. doi: 10.1038/s41598-017-03550-x.
This study aimed to compare the diagnostic capability of Pulsar perimetry (Pulsar) in pre-perimetric glaucoma (PPG) and early glaucoma (EG) with that of Flicker perimetry (Flicker) and spectral-domain optical conference tomography (SD-OCT). This prospective cross-sectional study included 25 eyes of 25 PPG patients, 35 eyes of 35 EG patients, and 42 eyes of 42 healthy participants. The diagnostic capability using the area under the curve (AUC) of the best parameter and agreement of detectability between structural and functional measurements were compared. For PPG patients, the AUC of Pulsar, Flicker, OCT-disc, and OCT-macular was 0.733, 0.663, 0.842, and 0.780, respectively. The AUC of Flicker was significantly lower than that of OCT-disc (p = 0.016). For EG patients, the AUC of Pulsar, Flicker, OCT-disc, and OCT-macular were 0.851, 0.869, 0.907, and 0.861, respectively. There was no significant difference in AUC among these methods. The agreement between structural and functional measurements expressed by kappa value ranged from -0.16 to 0.07 for PPG and from 0.01 to 0.25 for EG. Although the diagnostic capability of Pulsar in the PPG and EG groups was equal to that of Flicker and SD-OCT, the agreements between structural and functional measurements for both PPG and EG were poor.
本研究旨在比较脉冲式视野计(Pulsar)在前期青光眼(PPG)和早期青光眼(EG)中的诊断能力,与闪烁式视野计(Flicker)和光谱域光学相干断层扫描(SD-OCT)相比。本前瞻性横断面研究纳入了 25 例 PPG 患者的 25 只眼、35 例 EG 患者的 35 只眼和 42 例健康对照者的 42 只眼。比较了最佳参数的曲线下面积(AUC)和结构与功能测量之间的可检测性一致性的诊断能力。对于 PPG 患者,Pulsar、Flicker、OCT 盘和 OCT 黄斑的 AUC 分别为 0.733、0.663、0.842 和 0.780。Flicker 的 AUC 明显低于 OCT 盘(p=0.016)。对于 EG 患者,Pulsar、Flicker、OCT 盘和 OCT 黄斑的 AUC 分别为 0.851、0.869、0.907 和 0.861。这些方法之间的 AUC 没有显著差异。以 Kappa 值表示的结构和功能测量之间的一致性范围为 PPG 的-0.16 至 0.07,EG 的 0.01 至 0.25。虽然 Pulsar 在 PPG 和 EG 组中的诊断能力与 Flicker 和 SD-OCT 相当,但 PPG 和 EG 的结构和功能测量之间的一致性均较差。