VST Glaucoma Center, L V Prasad Eye Institute, Banjara Hills, Hyderabad, India.
Invest Ophthalmol Vis Sci. 2012 Feb 29;53(2):1058-65. doi: 10.1167/iovs.11-8463.
To evaluate the influence of a control group on the diagnostic accuracy of spectral-domain optical coherence tomography (SD-OCT) in early glaucoma.
In a diagnostic, case-control study, 119 eyes of 60 normal subjects with no findings suspicious for glaucoma (control cohort 1); 76 eyes of 41 subjects referred by general ophthalmologists as glaucoma suspects based on optic disc morphology, but found by glaucoma experts to be normal but with physiological variations in their optic nerves (control cohort 2); and 65 eyes of 46 early-glaucoma patients (cases) underwent imaging of the optic nerve head (ONH), retinal nerve fiber layer (RNFL), and ganglion cell complex (GCC) by SD-OCT.
Areas under the receiver operating characteristic curves (AUC) of ONH parameters discriminating glaucomatous eyes from normal eyes of control cohort 2 were significantly lesser (P < 0.001) than those discriminating glaucomatous eyes from normal eyes of control cohort 1. AUCs of RNFL parameters discriminating glaucomatous eyes from normal eyes of control cohorts 2 and 1 were comparable. Although the AUCs of GCC thickness parameters were comparable, AUCs of GCC focal and global loss volume in control cohort 2 (0.684 and 0.671. respectively) were significantly less (P < 0.05) than in control cohort 1 (0.881 and 0.841, respectively).
The effectiveness of most SD-OCT parameters in detecting glaucoma significantly decreased when evaluated against a clinically relevant control group with suspicious-looking optic nerves compared with that against a control group consisting of normal subjects with no findings suspicious for glaucoma.
评估对照组对早期青光眼光谱域光学相干断层扫描(SD-OCT)诊断准确性的影响。
在一项诊断性病例对照研究中,纳入 60 名无青光眼可疑表现的正常受试者的 119 只眼(对照组 1);纳入 41 名因视盘形态学被眼科医生怀疑为青光眼但经青光眼专家检查发现正常但视神经存在生理性改变的受试者的 76 只眼(对照组 2);纳入 65 名早期青光眼患者的 65 只眼(病例组)。对所有受试者的视神经头(ONH)、视网膜神经纤维层(RNFL)和节细胞复合体(GCC)进行 SD-OCT 成像。
与对照组 1 相比,对照组 2 中区分青光眼眼和正常眼的 ONH 参数的受试者工作特征曲线下面积(AUC)显著较小(P<0.001)。区分对照组 2 和对照组 1 中青光眼眼和正常眼的 RNFL 参数的 AUC 相似。虽然 GCC 厚度参数的 AUC 相似,但对照组 2 中 GCC 局灶性和总体丢失量的 AUC(分别为 0.684 和 0.671)显著较小(P<0.05),而对照组 1 中相应的 AUC(分别为 0.881 和 0.841)较大。
与对照组 1 相比,在评估疑似视神经的临床相关对照组时,大多数 SD-OCT 参数检测青光眼的有效性显著降低。