J Clin Virol. 2012 Apr;53(4):314-9. doi: 10.1016/j.jcv.2012.01.002. Epub 2012 Jan 30.
Rapid, high throughput extraction systems are needed to monitor viral infections in immunosuppressed patients.
Evaluate the performance of the MagNA Pure 96™ extraction system, and compare it to the COBAS Ampliprep™ for quantitative real-time PCR from whole blood samples.
Compare the MagNA Pure LC™, COBAS Ampliprep™ and MagNA Pure 96™ using ten-fold dilutions of blood samples containing cytomegalovirus. Evaluate analytical performances of the MagNA Pure 96™ from test samples containing cytomegalovirus. Evaluate clinical performances from 209 blood samples collected prospectively, extracted with the COBAS Ampliprep™ and the MagNA Pure 96™ systems and tested for cytomegalovirus, Epstein-Barr, BK and JC viruses.
All three extraction systems gave similar results with dilutions of a cytomegalovirus-positive sample. Analytical tests showed that the limit of detection was 500 copies/ml, specificity was 100%, with no cross-contamination. Quantification was linear from 3.0 to 6.0 log(10)copies/ml. Intra-assay variation was 8.3-0.9% and inter-assay variation 8.8-5.2%. Clinical specimens extracted with the MagNA Pure 96™ and COBAS Ampliprep™ instruments agreed well for cytomegalovirus (r=0.54; p=0.07), Epstein-Barr virus (0.69; p=0.0005) and BK virus (0.85; p=0.01). All 55 samples were negative for JC virus. Mean loads were similar for cytomegalovirus (0.17 log(10)copies/ml) and BK virus (-0.24 log(10)copies/ml) while that of Epstein-Barr virus was slightly lower (1.02 log(10)copies/ml).
The MagNA Pure 96™ instrument is an easy-to-use, reliable high throughput platform for extracting nucleic acid from clinical whole blood specimens.
需要快速、高通量的提取系统来监测免疫抑制患者的病毒感染。
评估 MagNA Pure 96 提取系统的性能,并将其与 COBAS Ampliprep 实时定量 PCR 从全血样本进行比较。
使用包含巨细胞病毒的血液样本十倍稀释液比较 MagNA Pure LC、COBAS Ampliprep 和 MagNA Pure 96。评估 MagNA Pure 96 从包含巨细胞病毒的测试样本中的分析性能。从 209 份前瞻性采集的血液样本中评估临床性能,这些样本用 COBAS Ampliprep 和 MagNA Pure 96 系统提取并检测巨细胞病毒、EB 病毒、BK 病毒和 JC 病毒。
三种提取系统对巨细胞病毒阳性样本的稀释液均给出相似的结果。分析测试表明,检测限为 500 拷贝/ml,特异性为 100%,无交叉污染。定量从 3.0 到 6.0 log(10)拷贝/ml 呈线性。批内变异为 8.3-0.9%,批间变异为 8.8-5.2%。用 MagNA Pure 96 和 COBAS Ampliprep 仪器提取的临床标本对巨细胞病毒(r=0.54;p=0.07)、EB 病毒(0.69;p=0.0005)和 BK 病毒(0.85;p=0.01)的检测结果一致。所有 55 份样本均为 JC 病毒阴性。巨细胞病毒(0.17 log(10)拷贝/ml)和 BK 病毒(-0.24 log(10)拷贝/ml)的平均负荷相似,而 EB 病毒的负荷略低(1.02 log(10)拷贝/ml)。
MagNA Pure 96 仪器是一种易于使用、可靠的高通量平台,可从临床全血标本中提取核酸。