J Am Dent Assoc. 2012 Feb;143(2):134-43. doi: 10.14219/jada.archive.2012.0123.
Nitrous oxide-oxygen sedation is used widely in dentistry because of its relative safety and efficacy. The most important safety consideration is the prevention of hypoxia. Safety features have been designed to prevent hypoxia by ensuring a minimal oxygen flow, thus limiting the amount of nitrous oxide that can be administered.
The authors review the 12 safety features used to ensure the safety and efficacy of nitrous oxide-oxygen sedation. They provide examples of safety feature failures, as well as steps to follow to help prevent negative outcomes.
Nitrous oxide-oxygen delivery systems typically are limited to a maximum of 70 percent nitrous oxide and 30 percent oxygen delivery, which ensures that the patient is receiving at least 9 percent more oxygen than found in ambient air. Other safety features stop the delivery of nitrous oxide if oxygen flow stops. The pin-index safety system prevents the accidental attachment of a nonoxygen tank to the oxygen attachment portal, and diameter-index systems help ensure that the appropriate gas flows through the appropriate tubing. Although these safety features are in place, dentists have reported incidents of hypoxia involving incorrect equipment installation or equipment damage. Practice Implications. If a safety feature failure is suspected during administration of nitrous oxide-oxygen sedation, the clinician should remove the face mask from the patient immediately.
由于一氧化二氮-氧气镇静具有相对安全性和有效性,因此在牙科中广泛使用。最重要的安全考虑因素是预防缺氧。安全功能旨在通过确保最小氧气流量来预防缺氧,从而限制可给予的一氧化二氮量。
作者回顾了用于确保一氧化二氮-氧气镇静安全性和有效性的 12 种安全功能。他们提供了安全功能故障的示例,以及帮助预防不良后果的步骤。
一氧化二氮-氧气输送系统通常限于最大 70%的一氧化二氮和 30%的氧气输送,这确保了患者接收的氧气比环境空气中的氧气多至少 9%。其他安全功能可在氧气流量停止时停止输送一氧化二氮。销索引安全系统可防止非氧气罐意外连接到氧气连接端口,而直径索引系统有助于确保适当的气体通过适当的管道流动。尽管存在这些安全功能,但牙医报告了涉及错误设备安装或设备损坏的缺氧事件。
实践意义。如果在给予一氧化二氮-氧气镇静时怀疑安全功能出现故障,临床医生应立即从患者身上取下面罩。