Licox(®)和劳美迪克(®)Neurovent-PTO脑组织氧监测仪测量准确性的实验比较

Experimental comparison of the measurement accuracy of the Licox(®) and Raumedic (®) Neurovent-PTO brain tissue oxygen monitors.

作者信息

Morgalla Matthias H, Haas R, Grözinger G, Thiel Christian, Thiel Karolin, Schuhmann Martin U, Schenk Martin

机构信息

Department of Neurosurgery, Eberhard Karls University Hospital, Tübingen, Germany.

出版信息

Acta Neurochir Suppl. 2012;114:169-72. doi: 10.1007/978-3-7091-0956-4_32.

Abstract

BACKGROUND

Only a few experimental reports are available on the direct comparison of Licox(®) and Raumedic(®)-Neurovent-PTO brain tissue oxygen pressure (P(br)O(2)) monitors. We compared the two systems regarding their measurement properties under experimental in vitro and in vivo conditions.

MATERIALS AND METHODS

Eight Licox(®) and Raumedic(®) Neurovent-PTO(®) sensors were tested for 10 min at 37°C, atmospheric pressure, at an oxygen content of 0% and 100% before and after the in vivo test. The same probes were implanted in German landrace pigs, which underwent hepatectomy. The mean P(br)O(2) values were recorded every minute. An O(2) challenge with inhalation of 100% O(2) for 10 min was performed 2 h post-abdominal surgery.

RESULTS

At 0% O(2) content values varied from 0.2 to 7 mmHg, at 100% O(2) content from 130 to 165 mmHg. No difference between probes was found. In vivo tests: Raumedic® showed higher P(br)O(2) values (mean +6.3 mmHg, p < 0.0001) compared with Licox®. During O(2) challenge, both probes responded similarly; however, Raumedic(®) had a 10% higher response amplitude (p < 0.005). After explantation there was again no difference between the two sensors.

CONCLUSION

Raumedic(®) sensors measured higher P(br)O(2) values. There was no significant difference regarding overall measurement of in vitro accuracy between the two probes, which proved to be robust when used consecutively for longer periods and in different environments.

摘要

背景

关于Licox®和劳美迪克(Raumedic®)-Neurovent-PTO脑组织氧分压(P(br)O(2))监测仪的直接比较,仅有少数实验报告。我们在体外和体内实验条件下比较了这两种系统的测量特性。

材料与方法

八个Licox®和劳美迪克Neurovent-PTO®传感器在37℃、大气压、氧含量为0%和100%的条件下,于体内测试前后各进行10分钟的测试。将相同的探头植入德国长白猪体内,这些猪接受了肝切除术。每分钟记录平均P(br)O(2)值。腹部手术后2小时,进行10分钟吸入100%氧气的氧激发试验。

结果

在氧含量为0%时,数值在0.2至7 mmHg之间变化;在氧含量为100%时,数值在从130至165 mmHg之间变化。未发现探头之间存在差异。体内测试:与Licox®相比,劳美迪克®显示出更高的P(br)O(2)值(平均高6.3 mmHg,p < 0.0001)。在氧激发试验期间,两种探头的反应相似;然而,劳美迪克®的反应幅度高10%(p < 0.005)。取出探头后,两种传感器之间再次没有差异。

结论

劳美迪克®传感器测量的P(br)O(2)值更高。两种探头在体外准确性的总体测量方面没有显著差异,当连续长时间在不同环境中使用时,证明它们是可靠的。