Department of Clinical Physiology, Nuclear Medicine and PET, Rigshospitalet, Copenhagen, University Hospital, Copenhagen, Denmark.
J Nucl Med. 2012 Mar;53(3):495-504. doi: 10.2967/jnumed.111.095240. Epub 2012 Feb 13.
Many authors have reported the importance of motion correction (MC) for PET. Patient motion during scanning disturbs kinetic analysis and degrades resolution. In addition, using misaligned transmission for attenuation and scatter correction may produce regional quantification bias in the reconstructed emission images. The purpose of this work was the development of quality control (QC) methods for MC procedures based on external motion tracking (EMT) for human scanning using an optical motion tracking system.
Two scans with minor motion and 5 with major motion (as reported by the optical motion tracking system) were selected from (18)F-FDG scans acquired on a PET scanner. The motion was measured as the maximum displacement of the markers attached to the subject's head and was considered to be major if larger than 4 mm and minor if less than 2 mm. After allowing a 40- to 60-min uptake time after tracer injection, we acquired a 6-min transmission scan, followed by a 40-min emission list-mode scan. Each emission list-mode dataset was divided into 8 frames of 5 min. The reconstructed time-framed images were aligned to a selected reference frame using either EMT or the AIR (automated image registration) software. The following 3 QC methods were used to evaluate the EMT and AIR MC: a method using the ratio between 2 regions of interest with gray matter voxels (GM) and white matter voxels (WM), called GM/WM; mutual information; and cross correlation.
The results of the 3 QC methods were in agreement with one another and with a visual subjective inspection of the image data. Before MC, the QC method measures varied significantly in scans with major motion and displayed limited variations on scans with minor motion. The variation was significantly reduced and measures improved after MC with AIR, whereas EMT MC performed less well.
The 3 presented QC methods produced similar results and are useful for evaluating tracer-independent external-tracking motion-correction methods for human brain scans.
许多作者已经报告了运动校正(MC)在 PET 中的重要性。患者在扫描过程中的运动干扰了动力学分析并降低了分辨率。此外,使用未对准的传输进行衰减和散射校正可能会在重建的发射图像中产生区域定量偏差。这项工作的目的是开发基于光学运动跟踪系统的人类扫描的外部运动跟踪(EMT)的 MC 程序的质量控制(QC)方法。
从 PET 扫描仪上获取的(18)F-FDG 扫描中选择了 2 次小运动扫描和 5 次大运动扫描(由光学运动跟踪系统报告)。运动被测量为附着在受试者头部的标记的最大位移,如果大于 4 毫米,则认为是大运动,小于 2 毫米则认为是小运动。在注射示踪剂后允许摄取时间为 40-60 分钟后,我们采集了 6 分钟的传输扫描,随后采集了 40 分钟的发射列表模式扫描。每个发射列表模式数据集被分为 5 分钟的 8 个帧。使用 EMT 或 AIR(自动图像配准)软件将重建的时间帧图像与选定的参考帧对齐。使用以下 3 种 QC 方法来评估 EMT 和 AIR MC:一种使用灰质体素(GM)和白质体素(WM)的 2 个感兴趣区域之间的比率的方法,称为 GM/WM;互信息;和交叉相关。
这 3 种 QC 方法的结果相互一致,并且与图像数据的视觉主观检查一致。在 MC 之前,QC 方法的测量在大运动扫描中差异很大,在小运动扫描中变化有限。在使用 AIR 进行 MC 后,变化明显减少,测量值提高,而 EMT MC 效果较差。
提出的 3 种 QC 方法产生了相似的结果,可用于评估用于人类大脑扫描的与示踪剂无关的外部跟踪运动校正方法。