Thoraxcenter, Erasmus MC, Rotterdam, The Netherlands.
EuroIntervention. 2012 Feb;7(10):1216-26. doi: 10.4244/EIJV7I10A194.
To assess the reproducibility of quantitative analysis of optical frequency domain imaging (OFDI) acquired at different pullback speeds (20, 30, 40 mm/sec), as well as the impact of cardiac motion artefact on three-dimensional (3D) reconstructions.
A total of 36 OFDI pullbacks were obtained pre- and post-stent implantation at the pullback speeds of 20, 30 and 40 mm/sec in non-diseased swine coronary arteries. The amount of x-ray contrast needed for blood clearance during OFDI imaging was recorded. Three-dimensional images of stented segments were rendered and artefacts on 3D images were assessed. Lumen areas (LA) were measured on each individual frame in pre- and post-stent pullbacks. The volume of contrast used with a pullback speed of 40 mm/sec was significantly smaller than with those of 30 and 20 mm/sec (10.8±1.8, 12.9±1.6, 15.9±2.6 ml, p<0.01, respectively). Three-dimensional reconstruction was feasible in all pullbacks. Faster pullback speeds resulted in a smaller number of artefacts. For quantitative measurement, a total of 7,426 frames were analysed. In non-stented vessels, LA derived from corresponding selected frames increased significantly with increasing pullback speeds (6.35±2.14 vs. 6.58±2.10 mm2 for 20 vs. 30 mm/sec [p<0.001], 6.36±2.13 vs. 6.75±2.09 mm2 for 20 vs. 40 mm/sec [p<0.001]), whereas in stented vessels there was no significant difference in mean LA between the three different pullback speeds (6.75±1.30 vs. 6.78±1.36 mm2 for 20 vs. 30 mm/sec [NS], 6.74±1.30 vs. 6.76±1.31 mm2 for 20 vs. 40 mm/sec [NS]).
Quantitative analysis of OFDI obtained at different pullback speeds in non-stented coronary arteries could potentially vary in LA measurement. OFDI with high-speed pullback allows quantitative analysis of stented vessels while reducing the amount of contrast and cardiac motion artefacts.
评估在不同回撤速度(20、30、40mm/sec)下获取的光频域成像(OFDI)定量分析的可重复性,以及心脏运动伪影对三维(3D)重建的影响。
在非病变猪冠状动脉中,以 20、30 和 40mm/sec 的速度进行支架植入前后共获得 36 次 OFDI 回撤。记录 OFDI 成像过程中清除血液所需的 X 射线造影剂用量。对支架节段的 3D 图像进行渲染,并评估 3D 图像上的伪影。在支架前和支架后每个单独的帧上测量管腔面积(LA)。以 40mm/sec 的回撤速度使用的造影剂体积明显小于 30 和 20mm/sec 的速度(分别为 10.8±1.8、12.9±1.6、15.9±2.6ml,p<0.01)。所有回撤均可行三维重建。更快的回撤速度导致的伪影数量更少。对于定量测量,共分析了 7426 个帧。在非支架血管中,随着回撤速度的增加,相应选择帧的 LA 显著增加(20 与 30mm/sec 相比,6.35±2.14 与 6.58±2.10mm2[ p<0.001],20 与 40mm/sec 相比,6.36±2.13 与 6.75±2.09mm2[ p<0.001]),而在支架血管中,三种不同回撤速度之间的平均 LA 无显著差异(20 与 30mm/sec 相比,6.75±1.30 与 6.78±1.36mm2[ NS],20 与 40mm/sec 相比,6.74±1.30 与 6.76±1.31mm2[ NS])。
在非支架冠状动脉中以不同回撤速度获取的 OFDI 定量分析可能会导致 LA 测量值发生变化。高速回撤的 OFDI 允许对支架血管进行定量分析,同时减少造影剂用量和心脏运动伪影。