评估一种数字化浊度仪在快速血浆反应素试验中用于梅毒血清学诊断的效果。

Evaluation of a digital flocculation reader for the rapid plasma reagin test for the serological diagnosis of syphilis.

机构信息

Division of STD Prevention, Centers for Disease Control and Prevention, Atlanta, GA 30333, USA.

出版信息

Sex Transm Dis. 2012 Mar;39(3):223-5. doi: 10.1097/OLQ.0b013e3182389ab9.

Abstract

We described the ASiManager-AT digital flocculation reader to demonstrate concordance between visual and digital readings of the rapid plasma reagin test for detection of antibodies in the serum of patients with syphilis. A qualitative and quantitative rapid plasma reagin was performed on each serum samples giving a concordance of 98.6% and 99.7%, respectively, for reactives and 100% for nonreactives.

摘要

我们描述了 ASiManager-AT 数字絮凝读数仪,以展示梅毒患者血清中抗体的快速血浆反应素试验的视觉读数和数字读数之间的一致性。对每份血清样本进行定性和定量快速血浆反应素检测,分别得到 98.6%和 99.7%的反应性一致性,以及 100%的非反应性一致性。