Ultrasound Med Biol. 2012 Apr;38(4):593-600. doi: 10.1016/j.ultrasmedbio.2011.12.019. Epub 2012 Feb 15.
The present study describes the presence of pseudoenhancement during contrast-enhanced ultrasound (CEUS) imaging of human carotid arteries and the reproduction of this pseudoenhancement in vitro. Seventy patients underwent bilateral CEUS examination of the carotid arteries using a Philips iU22 ultrasound system equipped with a L9-3 ultrasound probe and SonoVue microbubble contrast. During CEUS of the carotid arteries, we identified enhancement in close proximity to the far wall, parallel to the main lumen. The location of this enhancement does not correlate to the anatomical location of a parallel vessel. To corroborate the hypothesis that this is a pseudoenhancement artifact, the enhancement was recreated in a tissue-mimicking material phantom, using the same ultrasound system, settings and contrast agent as the patient study. The phantom study showed that pseudoenhancement may be present during vascular CEUS and that the degree of pseudoenhancement is influenced by the size and concentration of the microbubbles. During vascular CEUS, identification of the artifact is important to prevent misinterpretation of enhancement in and near the far wall.
本研究描述了在人颈动脉的对比增强超声(CEUS)成像中伪增强的存在,并在体外重现了这种伪增强。70 例患者使用配备 L9-3 超声探头和 SonoVue 微泡造影剂的飞利浦 iU22 超声系统进行双侧颈动脉 CEUS 检查。在颈动脉 CEUS 中,我们在远壁附近、与主腔平行的位置识别到增强。这种增强的位置与平行血管的解剖位置不相关。为了证实这是一种伪增强伪影的假设,我们使用与患者研究相同的超声系统、设置和造影剂,在组织模拟材料体模中重现了这种增强。体模研究表明,在血管 CEUS 中可能存在伪增强,伪增强的程度受微泡的大小和浓度的影响。在血管 CEUS 期间,识别伪影很重要,以防止对远壁及其附近的增强的错误解释。