Segmentation for navigated control was in the first generation very time consuming. In the present version (NCU 2.0) the risk structure is segmented (instead of the work space), this leads to an enormous decrease in preparation time. In additional, new safety functions were integrated. The segmentation feasibility was tested on patient data and proved to be successful. The automatic stop function was tested on petrous bone models and showed no damage to the facial nerve.
第一代用于导航控制的分割非常耗时。在当前版本(NCU 2.0)中,对风险结构进行了分割(而不是工作空间),这使得准备时间大幅减少。此外,还集成了新的安全功能。在患者数据上测试了分割的可行性,结果证明是成功的。在颞骨模型上测试了自动停止功能,结果显示对面神经没有损伤。