Institute of Neuroscience and Medicine, Forschungszentrum Jülich GmbH, Jülich, Germany.
IEEE Trans Med Imaging. 2012 Jul;31(7):1372-80. doi: 10.1109/TMI.2012.2188903. Epub 2012 Feb 24.
In hybrid magnetic resonance-positron emission tomography (MR-PET) studies with the Siemens 3T MR-BrainPET scanner an instantaneous reduction of the PET sensitivity was observed during execution of certain MR sequences. This interference was investigated in detail with custom-made as well as standard clinical MR sequences. The radio-frequency pulses, the switched gradient fields and the constant magnetic field were examined as the relevant parameters of the magnetic resonance imaging (MRI) system as well as the air temperature within the PET detectors. Our investigation comprised the analysis of the analog PET signals, the total count rates, the geometric distribution of the count rate reduction within the BrainPET detector as well as reconstructed images. The fast switching magnetic field gradients were identified to distort the analog PET detector signals. The measured count rate reduction was found to be less than 3%, but only up to 2% in the case of echo planar imaging sequences, as applied in functional MRI. For clinical sequences routinely used in hybrid MR-BrainPET measurements, a correction method has been designed, implemented, and evaluated .
在西门子 3T MR-BrainPET 扫描仪的混合磁共振-正电子发射断层扫描(MR-PET)研究中,在执行某些磁共振序列期间观察到 PET 灵敏度的瞬时降低。使用定制的以及标准的临床磁共振序列详细研究了这种干扰。检查了射频脉冲、切换梯度场和恒定磁场作为磁共振成像(MRI)系统的相关参数,以及 PET 探测器内的空气温度。我们的研究包括对模拟 PET 信号、总计数率、BrainPET 探测器内计数率降低的几何分布以及重建图像进行分析。快速切换的磁场梯度被确定为会扭曲模拟 PET 探测器信号。测量到的计数率降低小于 3%,但在应用于功能磁共振成像的回波平面成像序列中仅为 2%。对于混合 MR-BrainPET 测量中常规使用的临床序列,已经设计、实施和评估了一种校正方法。