Rasania Suraj P, Mountantonakis Stavros, Patel Vickas V
Department of Internal Medicine, Seton Hall University School of Heath & Medical Sciences, Trenton, New Jersey, USA.
Pacing Clin Electrophysiol. 2012 Sep;35(9):e267-71. doi: 10.1111/j.1540-8159.2012.03348.x. Epub 2012 Mar 4.
T-wave oversensing can be a serious problem that often results in inappropriate device therapy. We report here a patient with binge alcohol use who received multiple, inappropriate ICD shocks due to T-wave oversensing from repolarization changes induced by acute alcohol intoxication and no other relevant metabolic derangements. Following recovery from his alcohol intoxication a few days later, the T-wave amplitude decreased so the device no longer inappropriately sensed or delivered therapies. This case represents an uncommon, but reversible, cause of T-wave oversensing that should be considered before more aggressive measures are taken to correct the abnormality.
T波感知过度可能是一个严重问题,常常导致不适当的设备治疗。我们在此报告一名大量饮酒的患者,其因急性酒精中毒引起的复极改变导致T波感知过度,且无其他相关代谢紊乱,从而接受了多次不适当的植入式心律转复除颤器(ICD)电击。几天后,他从酒精中毒中恢复,T波振幅降低,因此该设备不再不适当感知或发放治疗。该病例代表了一种罕见但可逆的T波感知过度原因,在采取更积极措施纠正异常之前应予以考虑。