Sensors (Basel). 2009;9(1):118-30. doi: 10.3390/s90100118. Epub 2009 Jan 7.
The response time of thermal electrolytic Ag/AgCl reference electrodes is defined by the porous structure that limits the rate at which traces of any previous solution are diluted by any new solution environment. The electrode stabilisation time when transferred from one electrolyte to another has been shown to change when different structural designs to the conventional sphere of Ag/AgCl are used. Electrodes fabricated with cylindrical architectures of Ag/AgCl have shown improved stability and reach equilibrium faster than spherical electrodes. This work has positive implications for the accuracy and throughput of primary pH measurements made using the Harned cell.
热电解 Ag/AgCl 参比电极的响应时间取决于限制任何先前溶液痕迹被任何新溶液环境稀释的速率的多孔结构。当从一种电解质转移到另一种电解质时,已经证明当使用传统的 Ag/AgCl 球体的不同结构设计时,电极稳定时间会发生变化。用 Ag/AgCl 的圆柱形结构制造的电极显示出比球形电极更快的稳定性和更快的达到平衡。这项工作对使用哈恩细胞进行的主要 pH 测量的准确性和通量具有积极意义。