Analytical Science Division, National Physical Laboratory, Teddington, Middlesex TW11 0LW, UK.
Sensors (Basel). 2011;11(8):8072-84. doi: 10.3390/s110808072. Epub 2011 Aug 17.
A questionnaire was completed by fourteen world leading national metrology institutes to study the influence of several variables in the preparation of Ag/AgCl electrodes on the accuracy of Harned cell measurements of pH. The performance of each institute in the last decade has been assessed based on their results in eight key comparisons, organized by the Bureau International des Poids et Measures Consultative Committee for Amount of Substance, involving the measurement of pH of phosphate, phthalate, carbonate, borate and tetroxalate buffer solutions. The performance of each laboratory has been correlated to the results of the questionnaire to determine the critical parameters in the preparation of Ag/AgCl electrodes and their sensitivities with respect to the accuracy of pH measurement. This study reveals that the parameters most closely correlated to performance in comparisons are area of electrode wire exposed to the electrolyte, diameter and porosity of the Ag sphere prior to anodisation, amount of Ag converted to AgCl during anodisation, stability times employed for electrodes to reach equilibrium in solution prior to measurement, electrode rejection criteria employed and purity of reagents.
一份问卷由 14 家世界领先的国家计量研究院完成,旨在研究在制备 Ag/AgCl 电极时,几个变量对 Harned 细胞 pH 值测量精度的影响。根据他们在过去十年中在国际计量局物质数量咨询委员会组织的八次关键比对中的结果,评估了每个研究所的表现,这些比对涉及磷酸盐、邻苯二甲酸酯、碳酸盐、硼酸盐和四氧化二氮缓冲溶液的 pH 值测量。将每个实验室的表现与问卷的结果相关联,以确定在制备 Ag/AgCl 电极时的关键参数及其对 pH 值测量精度的敏感性。这项研究表明,与比对表现最密切相关的参数是暴露在电解质中的电极丝的面积、在阳极氧化之前 Ag 球的直径和孔隙率、在阳极氧化过程中转化为 AgCl 的 Ag 量、在测量之前电极在溶液中达到平衡所需的稳定时间、电极的拒绝标准和试剂的纯度。