Yonehara K, Shinohara M, Kanaya K
Hitachi Ltd., Ibaraki, Japan.
J Electron Microsc Tech. 1990 Nov;16(3):240-8. doi: 10.1002/jemt.1060160305.
Qualitative and quantitative x-ray energy dispersive spectroscopy is now used successfully to analyze many features and processes in inorganic samples. When applied to inorganic samples, however, the results are often less satisfactory due to problems of preparation of organic samples, difficulty of measuring x-rays from organic samples, damage of the sample by the electron beam, and other practical problems. In the present study we used a high voltage transmission electron microscope equipped with an energy dispersive x-ray spectrometer to examine accurate quantitative standardless analysis of thin sections of an organic sample, human dentin. Based on our experiments we found the important parameters for quantitative analysis were sample thickness and appropriate choice of model sample. Further, we show that the method of Cliff and Lorimer can be used with biological samples at 200 kV, and we show that quantitative analysis of human dentin can be carried out at 200 kV. Finally, we show that areas of human dentin can be differentiated by their morphological characteristics and x-ray analyses obtained in the transmission electron microscope.
定性和定量X射线能量色散光谱法目前已成功用于分析无机样品中的许多特征和过程。然而,当应用于无机样品时,由于有机样品制备问题、测量有机样品X射线的困难、电子束对样品的损伤以及其他实际问题,结果往往不太令人满意。在本研究中,我们使用配备能量色散X射线光谱仪的高压透射电子显微镜来检验对有机样品人牙本质薄片进行准确的定量无标样分析。基于我们的实验,我们发现定量分析的重要参数是样品厚度和模型样品的适当选择。此外,我们表明Cliff和Lorimer方法可在200 kV下用于生物样品,并且我们表明人牙本质的定量分析可在200 kV下进行。最后,我们表明人牙本质区域可以通过其形态特征和在透射电子显微镜中获得的X射线分析来区分。